Skip to the content

FDA Medical Device Recalls (openFDA)

TELEFLEX LLC: Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal. Intended for patients requiring endotracheal intubation

Recall number
Z-1010-2023
Recalling firm
TELEFLEX LLC
Product
Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal. Intended for patients requiring endotracheal intubation
Status
Open, Classified
Initiated
2022-12-08
Posted
2023-01-23
Terminated
not on file
Reason
Potential for pilot balloon non-inflation or cuff non-deflation.
Root cause
Process control
Action
Teleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/08/2022 by FedEx 2nd day mail. The notices explain the issues, risk, and requested the following: "Action list number 1 Medical facilities 1. We request that you immediately check your inventory for product within the scope of this FSCA. Users should cease use and distribution of affected product and immediately quarantine the affected product. 2. If you have impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and fax it to 1-855-419-8507, including Attn: Customer Service or using the email address recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you do not have impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and return the form to Teleflex at the contact details provided. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you. 2. We request that you immediately check your inventory for impacted product. Cease use and distribution of impacted product and immediately quarantine the affected product. You may then return all product in scope. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined in actions 1 and 2 of this Action List Number 2. Upon completion of your actions, please forward the completed Acknowledgement Form to the e-mail address below. Important - Please ensure you only list batch numbers in scope of this Field Safety Notice
Quantity
2,270 units
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WA, WI, WV and PR. The countries of Albania, Armenia, Austria, Azerbaijan, Bahamas, Barbados, Belgium, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Croatia, Cyprus, Czech, Denmark, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, Ireland, Italy Israel, Jordan, Kuwait, Lebanon, Liechtenstein, Malaysia, Mayotte, Mexico, Moldova, Morocco, Nepal, Netherlands, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russia, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, South Africa, South Korea, Sweden, Switzerland, Taiwan Thailand, Turkey, Ukraine, United Arab Emirates.
Lot, serial or model codes
a. Product Code: 170060, Batch Number: KME21E1411, UDI: (01)04026704319558(17)260428(10)KME21E1411; b. Product Code: 170070, Batch Number (UDI): KME21D1430 ((01)04026704319572(17)260328(10)KME21D1430), KME21E1290 ((01)04026704319572(17)260428(10)KME21E1290); c. Product Code: 170075, Batch Number: KME21D1785, UDI:(01)04026704319589(17)260328(10)KME21D1785; d. Product Code: 170080, Batch Number (UDI): KME21D1204 ((01)04026704319596(17)260328(10)KME21D1204), KME21E1287 ((01)04026704319596(17)260428(10)KME21E1287); e. Product Code: 170085, Batch Number: KME21E1296, UDI:(01)4026704319602(17)260428(10)KME21E1296; f. Product Code: 170090, Batch Number: KME21E1232, UDI: (01)4026704319619(17)260428(10)KME21E1232
510(k) numbers
["K780900"]
PMA numbers
not on file
Product code
BTR
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register