Solta Medical, Inc.: Fraxel re:store Dual Laser System----a laser surgical instrument for use in general and plastic surgery and in dermatology. Model Numbers: MC-SYS-SR1500-D-US; MC-SYS-SR1500-D-US-LOANER; MC-SYS-SR1500-D-UPG-US; MC-SYS-SR 1500-0-1 NTL; MC-SYS-SR1500-D-I-LOANER; MC-SYS-SR1500-D-UPG-INTL; Product is manufactured and distributed by Solta Medical, Inc., Hayward, CA The Fraxel re:store Dual Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue.

Recall number
Z-1010-2011
Recalling firm
Solta Medical, Inc.
Product
Fraxel re:store Dual Laser System----a laser surgical instrument for use in general and plastic surgery and in dermatology. Model Numbers: MC-SYS-SR1500-D-US; MC-SYS-SR1500-D-US-LOANER; MC-SYS-SR1500-D-UPG-US; MC-SYS-SR 1500-0-1 NTL; MC-SYS-SR1500-D-I-LOANER; MC-SYS-SR1500-D-UPG-INTL; Product is manufactured and distributed by Solta Medical, Inc., Hayward, CA The Fraxel re:store Dual Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue.
Status
Terminated
Initiated
2010-12-15
Posted
2011-02-01
Terminated
2011-11-14
Reason
The product has the potential to cause inadvertent laser firing which may lead to patient/operator injury if the problem recurs.
Root cause
Software design
Action
The firm, Solta Medical, issued three "DEVICE CORRECTION" letters, one dated December 15, 2010, the second one date December 17, 2010 and the third one dated December 20, 2010, to all affected customers. The letters described the product, problem and action to be taken by the customers. The customers were instructed to follow the instructions carefully, sign and date the bottom of the instruction form upon completion of the update and promptly return the signed form and USB stick in the enclosed, return-addressed envelope (Product Support, Solta Medical, 25881 Industrial Blvd., Hayward, CA 94545) Note: Solta Medical sales personnel will assist with the software update process so, if your local representative has already updated your system, you do not need to do anything further. If you have any questions, please call Solta Product Support at (510) 259-7291.
Quantity
352 units
Distribution
Worldwide Distribution: USA and countries including: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Egypt, Germany, Greece, Hong Kong, India, Italy, Japan, Korea, Kuwait, Mexico, Netherlands, New Zealand, Pakistan, Philippines, Poland, Russia, Singapore, South Africa, Spain, Thailand, Turkey, Ukraine, United Kingdom, and Vietnam
Lot, serial or model codes
Serial Numbers: F1196J, J0007, J0010-J0058, J0060-J0203, J0205, J0207-J0209, J0211-J0337, J0339-J0359, J0361-J0414, J0417-J0419, J0426
510(k) numbers
["K091420","K101490"]
PMA numbers
not on file
Product code
ONG
Firm FEI
3003079710
Firm city
Hayward
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register