FDA Medical Device Recalls (openFDA)
Roche Diagnostics Operations, Inc.: The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and electrochemiluminescence detection. Patient serum sample is incubated with pretreatment reagent PT1 and the reconstituted PT2 containing a pre-formed immunocomplex of solubilized porcine TSH receptor (pTSHR) and biotinylated anti-pTSHR mouse monoclonal antibody.
- Recall number
- Z-1009-2026
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Product
- The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and electrochemiluminescence detection. Patient serum sample is incubated with pretreatment reagent PT1 and the reconstituted PT2 containing a pre-formed immunocomplex of solubilized porcine TSH receptor (pTSHR) and biotinylated anti-pTSHR mouse monoclonal antibody.
- Status
- Open, Classified
- Initiated
- 2025-11-21
- Posted
- 2026-01-02
- Terminated
- not on file
- Reason
- Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
- Root cause
- Process control
- Action
- Firm notified customers on November 21, 2025 via Urgent Medical Device Correction notices. Customers using the cobas e 402/801 and cobas e411 were instructed to immediately discontinue using the impacted lots outlined in the UMDC and switch to the new unaffected lot numbers listed in the communication.
- Quantity
- 1300 cassettes
- Distribution
- US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI, MN, NC, NJ, NM, NY, OH, PA, PR, TN, TX, UT, WA, WV.
- Lot, serial or model codes
- 08496633190 lot 840183 exp. 02/26 08496633190 lot 874011 exp. 10/26 08496609190 lot 840177 exp. 05/26
- 510(k) numbers
- ["K193313"]
- PMA numbers
- not on file
- Product code
- JZO
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28