Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Molecular, Inc.: Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.

Recall number
Z-1009-2023
Recalling firm
Abbott Molecular, Inc.
Product
Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.
Status
Open, Classified
Initiated
2022-12-09
Posted
2023-01-20
Terminated
not on file
Reason
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
Root cause
Under Investigation by firm
Action
On December 09, 2022, Abbott Molecular issued a Field Correction Recall/Urgent Field Safety Notice to all customers who have the Alinity m System by FedEx priority overnight express. The letter explained the problems and provided the following information: "An assessment of your Alinity m System associated with the above listed potential issues will be completed remotely if your system is currently connected to AbbottLink or alternatively via onsite inspection by an Abbott representative. If necessary, an Abbott representative will schedule time to corrected the affected ADU9s) on your Alinity m System(s). In the interim, please continue to follow your laboratory protocols for any suspected incorrect results. Following the assessment and any necessary service, an additional communication will follow. If it is determined that your instrument was impacted, please assess the impact to your laboratory in reviewing previously generated test results. This recall is to be carried out at the user/customer level. If this product has been further distributed by your facility, please notify any additional impacted customers. Please review this information with laboratory personnel and retain this communication for future reference. If you have any questions regarding the communication, please contact your local Abbott representative, or call Abbott Technical Support at 1-800-553-7042 Option 2."
Quantity
918 units
Distribution
x
Lot, serial or model codes
Model Number: 08N53-002
510(k) numbers
not on file
PMA numbers
["P190025"]
Product code
OOI
Firm FEI
3005248192
Firm city
Des Plaines
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register