Covidien LP: AutoSuture SURGIPRO Polyproylene Mesh Clear 3" x 5" Sterile REF SPM 35 Surgipro Flat Sheet mesh is a non-absorbable, polypropylene surgical mesh exhibiting high burst strength and tensile strength. Surgipro comes in multiple designs, including multi-filament and open weave. All designs are knitted in such a fashion as to provide bi-directional elasticity, while allowing the mesh to be cut to any shape without unraveling. Mesh is intended to assist in the repair and/or reinforcement of hernia and other fascial defects requiring the additional support of a non-absorbable mesh during and after wound healing.

Recall number
Z-1009-2011
Recalling firm
Covidien LP
Product
AutoSuture SURGIPRO Polyproylene Mesh Clear 3" x 5" Sterile REF SPM 35 Surgipro Flat Sheet mesh is a non-absorbable, polypropylene surgical mesh exhibiting high burst strength and tensile strength. Surgipro comes in multiple designs, including multi-filament and open weave. All designs are knitted in such a fashion as to provide bi-directional elasticity, while allowing the mesh to be cut to any shape without unraveling. Mesh is intended to assist in the repair and/or reinforcement of hernia and other fascial defects requiring the additional support of a non-absorbable mesh during and after wound healing.
Status
Terminated
Initiated
2010-12-15
Posted
2011-02-01
Terminated
2013-01-10
Reason
Sterility compromised due to breach in sterile barrier
Root cause
Packaging process control
Action
Covidien notified customers via letter dated November 15, 2010. The letter identified the product, the problem, and action to be taken by the customer. Customers were instructed to examine inventory and return the affected lot to Field Returns Department, 195 McDermott Road, North Haven, CT 06473. For questions regarding this recall call 203-492-8091.
Quantity
132 units
Distribution
Worldwide Distribution - USA including CA, CT, FL, HI, IL, IN, KS, MA, MD, MI, MN, NC, NJ, NY, OH, PA, SC, TX, and WI and the countries of Canada and Italy
Lot, serial or model codes
Lot Number: A0K0490 Exp Date: 2015-10
510(k) numbers
["K915526"]
PMA numbers
not on file
Product code
FTL
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register