Carl Zeiss Meditec AG: ZEISS miLOOP Lens Fragmentation Device REF 303071-9090-000. Used during cataract surgery.

Recall number
Z-1007-2022
Recalling firm
Carl Zeiss Meditec AG
Product
ZEISS miLOOP Lens Fragmentation Device REF 303071-9090-000. Used during cataract surgery.
Status
Open, Classified
Initiated
2022-03-16
Posted
not on file
Terminated
not on file
Reason
High friction of the slider can cause the device to stick, or not move as intended.
Root cause
Under Investigation by firm
Action
On 03/16/2022, the Recalling firm called U.S. Customers to inform them of that is has been determined that single lot (Lot FG21082410) of the ZEISS miLoop may have been provided with potentially too high friction of the slider. A follow-up email was sent on 03/22/2022 that contained an "Urgent Medical Device Recall" letter. Customer are asked to discontinue the use and not to further distribute the specific product lot, and to fill out the attached acknowledgement form to dl.med-qualitycti.us@zeiss.com. Customer are asked to return all devices from lot FG21082410 to Carl Zeiss Meditec AG. For question or assistance, contact: Scott Smith, Director of Sales at (704) 906-3287 or email scott.smith@zeiss.com
Quantity
1225 devices
Distribution
U.S.: AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OR, PA, Puerto Rico, SC, TN, TX. UT, VA, WA, and WV O.U.S.: Not provided
Lot, serial or model codes
Model: 303071-9090-000 Catalog Number: FG-50608 UDI: (01)04049539104496(11)201102(17)231031(10)FG21082410 Lot Number: FG21082410
510(k) numbers
not on file
PMA numbers
not on file
Product code
HNQ
Firm FEI
3003481798
Firm city
Oberkochen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register