Arrow International Inc: Multi-Lumen PI CVC Kit and PSI Kit: 3-L 7 FR X 8 product code: ASK-45703-POSU lot #23F13L0166, exp. 31-Jul-2015: Product Usage: Multi-Lumen PI CVC Kit: An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patients vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials. PSI Kit: A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Recall number
Z-1005-2015
Recalling firm
Arrow International Inc
Product
Multi-Lumen PI CVC Kit and PSI Kit: 3-L 7 FR X 8 product code: ASK-45703-POSU lot #23F13L0166, exp. 31-Jul-2015: Product Usage: Multi-Lumen PI CVC Kit: An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patients vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials. PSI Kit: A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Status
Terminated
Initiated
2014-12-16
Posted
2015-01-22
Terminated
2015-08-10
Reason
the Instructions for Use (IFU) were inadvertently removed from the kit.
Root cause
Labeling False and Misleading
Action
An Urgent Medical Device Corrective Action Notification letter, dated December 16, 2014, was sent to Owens & Minor, 1605 Worldwide Blvd., Hebron, KY 41048-8653, instructing them to immediately discontinue distribution and quarantine any products, use customer letter and Corrective Action Acknowledgement Form templates to communicate this corrective action to any of thier customers that have received product included within the scope of the corrective action, and to have their customers return a completed Corrective Action Acknowledgement Form outlining their remaining inventory of the affected product codes and lot numbers for which replacement IFUs is needed.
Quantity
420
Distribution
In the state of KY
Lot, serial or model codes
Multi-Lumen PI CVC Kit: 3-L 7 FR X 8 product code: ASK-45703-POSU lot #23F13L0166, exp. 31-Jul-2015  PSI Kit: 8.5 FR X 4-1/8 (10 CM), product code ASK-09803-OSU2 , lot #23F14C1014, exp. 31-Aug-2015:
510(k) numbers
["K071538"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register