FDA Medical Device Recalls (openFDA)
Ossur Americas: Ossur, LIL Angel Vest, Part/ Description: 545100K/ LIL ANGEL VEST P2 KODEL; 545100L/ LIL ANGEL VEST P2 LAMBSWOOL; 545100S/ LIL ANGEL VEST P2 SORBATEX; 545150K/ LIL ANGEL VEST P3 KODEL; 545150L/ LIL ANGEL VEST P3 LAMBSWOOL; 545150S/ LIL ANGEL VEST P3 SORBATEX
- Recall number
- Z-1004-2019
- Recalling firm
- Ossur Americas
- Product
- Ossur, LIL Angel Vest, Part/ Description: 545100K/ LIL ANGEL VEST P2 KODEL; 545100L/ LIL ANGEL VEST P2 LAMBSWOOL; 545100S/ LIL ANGEL VEST P2 SORBATEX; 545150K/ LIL ANGEL VEST P3 KODEL; 545150L/ LIL ANGEL VEST P3 LAMBSWOOL; 545150S/ LIL ANGEL VEST P3 SORBATEX
- Status
- Terminated
- Initiated
- 2018-06-21
- Posted
- not on file
- Terminated
- 2020-09-04
- Reason
- Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
- Root cause
- Labeling design
- Action
- On 06/21/2018, Urgent Field Safety Notices were mailed to customers via UPS Next-Day Air requiring a signature of acceptance from the customer. Customers were advised the following: 1. Please pass this notice to those who need to be aware within your organization or to any department areas where the potentially affected devices will be used or transferred. 2. Maintain awareness on this notice for an appropriate period. 3. If you would like to have the product replaced please contact your customer service or sales representative and new product will be provided. Customers can exchange the affected product, if they desire to do so. Customers were asked to distribute the notices to those who need to be aware within their organization or to any department areas where the potentially affected devices were used or transferred. On 01/14/2019, Urgent Medical Device Recall letters were sent via express mail requiring a signature of acceptance from the customer. Customers were advised: Please examine your inventory, quarantine products subject to the recall and contact customer service for a return authorization to return the product. In addition: 1. Please confirm receipt of this recall notice by calling customer service (800-233-6263) or visit https://safetyalert.ossur.com 2. Please pass this notice to those who need to be aware within your organization and to any department areas where the potentially affected devices will be used or transferred (e.g. MR Imaging, Surgical, Emergency, Material Management, Patient) 3. Maintain awareness on this notice for an appropriate period. 4. If you have further distributed this product, please identify your customers and notify them at once of this product alert. We recommend that you include a copy of this product recall. 5. If any of your patients is currently wearing the product, necessary arrangements should be made to ensure the patient is not placed in MR Imaging.
- Quantity
- 45
- Distribution
- U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
- Lot, serial or model codes
- All lots of the products manufactured from 06/30/2009 - 12/13/2018
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KQZ
- Firm FEI
- 1000519909
- Firm city
- Foothill Ranch
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.