Ortho-Clinical Diagnostics: VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS System family member, 2) 5600 Integrated System 1) VITROS 4600 - in vitro quantitative measurement of a variety of analytes, 2) VITROS 5600 - in vitro quantitative, semi-quantitative, and qualitative measurement of analytes
- Recall number
- Z-1004-2015
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS System family member, 2) 5600 Integrated System 1) VITROS 4600 - in vitro quantitative measurement of a variety of analytes, 2) VITROS 5600 - in vitro quantitative, semi-quantitative, and qualitative measurement of analytes
- Status
- Terminated
- Initiated
- 2014-09-29
- Posted
- 2015-01-20
- Terminated
- 2017-04-03
- Reason
- Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
- Root cause
- Software Manufacturing/Software Deployment
- Action
- Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated September 29, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2. The resolution to this software anomaly will be in a future version of software. Until the new software is installed on their VITROS System, customers were instructed not to use the calibrator barcode labels for VITROS Calibrator Kit 2. Customers were advised that you manually program the calibration. Customers with questions were instructed to contact Customer Technical Services at 1-800-421-3311.
- Quantity
- 12 units
- Distribution
- Worldwide Distribution - USA including CO, NC, NJ, NY, PA, MI, WV and Internationally to Canada, Japan and Spain.
- Lot, serial or model codes
- System Product Code: 1) 6802445, 2) 6802413
- 510(k) numbers
- ["K031924","K063144"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28