Skip to the content

FDA Medical Device Recalls (openFDA)

Veridex, LLC: CellTracks Analyzer II. A semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immuno-magnetically selected and aligned. The system is for in-vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

Recall number
Z-1004-2012
Recalling firm
Veridex, LLC
Product
CellTracks Analyzer II. A semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immuno-magnetically selected and aligned. The system is for in-vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.
Status
Terminated
Initiated
2011-08-17
Posted
2012-02-13
Terminated
2015-06-17
Reason
Control cell images may i intermix with Patient cell images while changing between the Control review screen and Patient review screen on the Cell tracks Analyzer II.
Root cause
Software design
Action
Veridex, LLC sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letter dated August 17, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter provides procedure sheets with step by step directions for customers to follow. Additionally, a Confirmation of Receipt form was enclosed for customers to complete and return to the firm. Contact Customer Technical Services at 1-877-837-4339 for questions regarding this notice.
Quantity
37 Units
Distribution
Worldwide Distribution.
Lot, serial or model codes
K050145 (March 15, 2005); K060110 (March 16, 2006) Product code 9555
510(k) numbers
["K050145","K060110","K113181"]
PMA numbers
not on file
Product code
NQI
Firm FEI
3004619490
Firm city
Raritan
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register