Optovue Inc.: Optovue iVue optical coherence tomography system, computer controlled opthalmic imaging system, manufactured by Optovue, Fremont, CA Computer controlled opthalmic imaging system. Device scans the patient's eye using low coherence interferometer to measure the reflectivity of the retinal tissue. The cross sectional retinal tissue structure is composed of a sequence of A-scans. It has a traditional patient and instrument interface. The computer has a graphic user interface for acquiring and analyzing the image. The iVue offers three scans: Retina, Glaucoma and Cornea. For the cornea scan, a lens must be attached to the front of the device for proper scanning. This is the Cornea Anterior Module (CAM).

Recall number
Z-1004-2011
Recalling firm
Optovue Inc.
Product
Optovue iVue optical coherence tomography system, computer controlled opthalmic imaging system, manufactured by Optovue, Fremont, CA Computer controlled opthalmic imaging system. Device scans the patient's eye using low coherence interferometer to measure the reflectivity of the retinal tissue. The cross sectional retinal tissue structure is composed of a sequence of A-scans. It has a traditional patient and instrument interface. The computer has a graphic user interface for acquiring and analyzing the image. The iVue offers three scans: Retina, Glaucoma and Cornea. For the cornea scan, a lens must be attached to the front of the device for proper scanning. This is the Cornea Anterior Module (CAM).
Status
Terminated
Initiated
2010-07-20
Posted
2011-02-01
Terminated
2011-02-08
Reason
Use of the CAM attachment may result in a positioning anomaly and mistreatment in the eyes of the patient.
Root cause
Software design (manufacturing process)
Action
Optovue Inc. sent a Field Modification/Correction Notice dated July 20, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were informed that they could continue to use the iVue until the correction has been made. Customers were instructed to review any existing iVue patient data to ensure that this situation and an improper designation of eye assignment has not occurred. Upon completion of the updates, customers were to sign the attached form indicating they acknowledge that their iVue has been updated and they understand that they must review existing data. For any questions regarding this recall call 1-866-344-8948 or by mail at 45531 Northport Loop W, Fremont, CA 94538.
Quantity
22 devices in the US.
Distribution
Nationwide
Lot, serial or model codes
Serial numbers: 20305, 20307, 20340, 20335, 20306, 20313, 20339, 20337, 20338, 20341, 20311, 20271, 20343, 20312, 20309, 20310, 20308, 20336, 20344, 20349, 20346, 20347
510(k) numbers
["K091404"]
PMA numbers
not on file
Product code
NQQ
Firm FEI
3005950902
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register