Boston Scientific Corporation: SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.
- Recall number
- Z-1003-2022
- Recalling firm
- Boston Scientific Corporation
- Product
- SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.
- Status
- Open, Classified
- Initiated
- 2022-02-24
- Posted
- not on file
- Terminated
- not on file
- Reason
- Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
- Root cause
- Labeling design
- Action
- An URGENT MEDICAL DEVICE PRODUCT ADVISORY dated 2/24/22 was sent to customers. Inadvertent placement of SpaceOAR gel into a blood vessel and subsequent migration of the hydrogel outside of the pelvis, led to embolism. Boston Scientific will incorporate best practices regarding hydrogel placement during all SpaceOAR procedures into the IFU. The updated IFUs will be packaged with SpaceOAR and SpaceOAR Vue Systems after regulatory approvals are obtained for the IFU updates. This Product Advisory contains the recommended IFU updates, located in Appendix 1, intended to: -Raise awareness of embolism as an existing adverse event. -Add new Warnings and Precautions that describe technique recommendations for the placement of SpaceOAR and SpaceOAR Vue to reduce the unintentional placement of hydrogel. - Add procedural instructions to reduce the unintentional placement of hydrogel in a blood vessel and steps to verify placement of the hydrogel following implantation. NO product is being recalled and you are NOT required to return product to Boston Scientific. As per the IFU, SpaceOAR and SpaceOAR Vue System procedures should only be performed by physicians who have received appropriate training for proper hydrogel spacer deployment technique. The Acknowledgement Form must be completed and returned even if you do not have any affected units. Return the Acknowledgement Form by Email: BSCFieldActionCenter@bsci.com or Fax to: Field Action Center 1-763-415-7708
- Quantity
- 23,368 units
- Distribution
- Worldwide - US Nationwide distribution.
- Lot, serial or model codes
- UPN SV-1010, GTIN 00850009803009, Batch Numbers: 26383209, 26383210, 26517604, 26861584, 26861585, 26891612, 26891613, 27083784, 27190242, 27210560, 27257801, 27257802, 27455966, 27455967, 27496959, 27497140, 27670956, 27708041, 27708042, 27708043, 27930863, 27930864, 27941129, 27941130, 27949315, 28094077, 28247077, 28247078, 28358778, 28377018, 28377019, 28464794, 28464795, 28496461, 28496462, 28578758, 28578759, 28608231, 28608232; UPN SV-2101, GTIN 00864661000140, Batch Numbers: 26345481, 26376961, 26376962, 26376963, 26376964, 26380456, 26380457, 26384055, 26384056, 26408849, 26485974, 26485975, 26485977, 26486586, 26486587, 26486588, 26493461, 26517606, 26517607, 26529719, 26567283, 26567284, 26567285, 26567286, 26567287, 26574373, 26574374, 26574378, 26574379, 26574441, 26611593, 26611594, 26611595, 26652286, 26652287, 26691824, 26691826, 26691827, 26691828, 26705563, 26705564, 26705565, 26705566, 26728564, 26728565, 26743769, 26743770, 26743771, 26743772, 26745057, 26770161, 26770162, 26770163, 26770164, 26770165, 26861586, 26880769, 26880770, 26880771, 26880772, 26898042, 26898043, 26898044, 26898045, 26899812, 26899813, 26906603, 26906604, 26929712, 26929713, 26929714, 26929715, 26929716, 26929717, 26929719, 26929860, 26929861, 26957514, 26957515, 26957516, 26957517, 26969537, 26969538, 26969539, 26970440, 26970441, 26970442, 27010832, 27040235, 27040236, 27060422, 27060423, 27060424, 27060425, 27060426, 27078872, 27078873, 27084457, 27084458, 27084459, 27084960, 27100768, 27100769, 27100770, 27123402, 27123404, 27140485, 27140486, 27163487, 27163488, 27180605, 27190243, 27190244, 27190245, 27190686, 27193837, 27193838, 27193839, 27194460, 27194461, 27197622, 27197623, 27214381, 27214382, 27214383, 27224419, 27257804, 27257805, 27259062, 27259063, 27259064, 27259065, 27268186, 27268187, 27283462, 27283463, 27283464, 27283465, 27283466, 27317224, 27317225, 27317226, 27329273, 27329274, 27329275, 27358004, 27358005, 27361380, 27361381, 27382203, 27382204, 27416003, 27416004, 27416005, 27416006, 27416007, 27444898, 27444899, 27445360, 27447905, 27469811, 27469812, 27469813, 27475230, 27496957, 27496958, 27512052, 27512053, 27512054, 27512055, 27517685, 27518868, 27518869, 27528492, 27531152, 27531153, 27531154, 27531155, 27531156, 27531157, 27601494, 27601495, 27657444, 27657445, 27657446, 27670942, 27670945, 27670946, 27670949, 27705891, 27720117, 27720118, 27720119, 27745215, 27745216, 27756475, 27769338, 27769339, 27769860, 27769861, 27769862, 27769863, 27797083, 27797084, 27797085, 27797086, 27797087, 28094078, 28094079, 28094320, 28103330, 28104177, 28104178, 28104179, 28104420, 28110684, 28110685, 28110686, 28110687, 28116634, 28116635, 28138219, 28138440, 28146630, 28146631, 28169753, 28169754, 28169755, 28179669, 28179670, 28179671, 28202160, 28237387, 28237388, 28237389, 28247079, 28247160, 28247161, 28247162, 28247269, 28247270, 28247271, 28247272, 28254386, 28254387, 28254388, 28265943, 28265944, 28265945, 28265946, 28265947, 28272556, 28272557, 28272558, 28272559, 28272560, 28287355, 28287356, 28294718, 28294719, 28311111, 28311112, 28323414, 28323415, 28323416, 28323417, 28335613, 28335614, 28350366, 28350367, 28350368, 28350369, 28364006, 28364007, 28364008, 28371227, 28371228, 28371229, 28389635, 28389636, 28412353, 28419102, 28419103, 28432892, 28432893, 28473289, 28473290, 28493777, 28493778, 28499505, 28499506, 28499507, 28555843, 28565355, 28565356, 28565357, 28587389, 28587390, 28587391, 28587392, 28587393, 28587394, 28605867, 28606252, 28611099, 28612440, 28612441, 28612442, 28613189, 28635760, 28635761, 28635762, 28635763, 28654059, 28654440, 28654441, 28663263, 28663264, 28663265, 28668946, 28668947, 28668948, 28683114, 28683115, 28686271, 28686272, 28686273, 28694687, 28707785, 28709642, 28709643, 28714572, 28723144, 28734763, 28734764, 28758266, 28766992, 28766993, 28766994, 28766995, 28766996, 28811555, 28811556, 28811558, 28811560, 28819972, 28819973, 28821692, 28821693, 28822526, 28822527, 28828281, 28828282, 28836638, 28836639, 28838125
- 510(k) numbers
- ["K182971"]
- PMA numbers
- not on file
- Product code
- OVB
- Firm FEI
- 3005099803
- Firm city
- Marlborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28