FDA Medical Device Recalls (openFDA)
DeRoyal Processing Center: Custom surgical kits: DeRoyal(R) TRAUMA CRANIOTOMY PACK, REF 89-7639.01, STERILE EO, Manufactured by: DeRoyal Industries, Inc., Powell, TN 37849 DeRoyal(R) TRAUMA CRANIOTOMY PACK PGYBK, REF 89-7639.01, STERILE EO, Manufactured by: DeRoyal Industries, Inc., Powell, TN 37849 DeRoyal(R) TRAUMA CRANIOTOMY PACK PGYBK, REF 89-7639.02, STERILE EO, Manufactured by: DeRoyal Industries, Inc., Powell, TN 37849 surgical kits
- Recall number
- Z-1003-2012
- Recalling firm
- DeRoyal Processing Center
- Product
- Custom surgical kits: DeRoyal(R) TRAUMA CRANIOTOMY PACK, REF 89-7639.01, STERILE EO, Manufactured by: DeRoyal Industries, Inc., Powell, TN 37849 DeRoyal(R) TRAUMA CRANIOTOMY PACK PGYBK, REF 89-7639.01, STERILE EO, Manufactured by: DeRoyal Industries, Inc., Powell, TN 37849 DeRoyal(R) TRAUMA CRANIOTOMY PACK PGYBK, REF 89-7639.02, STERILE EO, Manufactured by: DeRoyal Industries, Inc., Powell, TN 37849 surgical kits
- Status
- Terminated
- Initiated
- 2012-01-10
- Posted
- 2012-02-13
- Terminated
- 2012-11-05
- Reason
- The firm distributed surgical kits which contained DermaHook Neuro Elastic Retractors which were subsequently recalled by Teleflex.
- Root cause
- Nonconforming Material/Component
- Action
- DeRoyal Industries sent a Customer letter dated January 6, 2012, to the affected customer. The letter identified the product the problem and the action needed to be taken by the customer. An email was also sent on January 10, 2012, to the affected customer. The customer was requested to follow these instructions. 1) Identify any affected inventory using the product and lot numbers on Attachment 1. 2) Attach the provided warning label to each affected kit and tray. 3) Forward this recall to any of your end users that may have affected product. 4) Complete and return Attachment 1 by fax (865-362-3716) or email (recalls@deroyal.com) within two weeks of the date of this letter, even if you no longer have affected inventory. If you have questions or need assistance with the recall, including additional labels, please call (865) 362-1037 . We apologize for any inconvenience this may caused you.
- Quantity
- 214 units
- Distribution
- (USA) including Indiana
- Lot, serial or model codes
- REF 89-7639.01 - Lot Number 24579045 REF 89-7639.01 - Lot Number 24664258 REF 89-7639.02 - Lot Numbers: 24828048, 25598985, 25359212, 25399676
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRP
- Firm FEI
- 1000208624
- Firm city
- New Tazewell
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28