FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corp: Terumo Perfusion System 8000 Roller Pump, 115V; Model 8000; Catalog No. 16402.
- Recall number
- Z-1002-2007
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Product
- Terumo Perfusion System 8000 Roller Pump, 115V; Model 8000; Catalog No. 16402.
- Status
- Terminated
- Initiated
- 2004-07-01
- Posted
- 2007-07-07
- Terminated
- 2007-12-12
- Reason
- Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being non-functional.
- Root cause
- Other
- Action
- As the complaints were received (between 07/04 and 01/06), the firm visited the customer filing the complaint and repaired the pumps demonstrating the problem. The firm will correct other units as complaints are received. The firm modified its recall strategy and issued a Device Correction Letter dated 6/22/07 notifying all users of the problem
- Quantity
- 728 pumps of all models
- Distribution
- Worldwide: USA and Bangladesh, Belgium, Bolivia, Canada, Chile, Colombia, Costa Rica, Egypt, Equador, Hong Kong, India, Indonesia, Japan, Latin America, Malaysia, Nicaragua, Philippines, Saudi Arabia, Taipe, Taiwan, United Arab Emirates, Uruguay, and Viet Nam.
- Lot, serial or model codes
- Serial numbers 10031, 10033, 10137, 10138, 10614, 10663, 10730, 11131, 11281, 11283, 11361, 11548, 11630, 11751, 12110, 12604, 12686, 12938, 13129, 13719, 13826, 13830, 14712, 15090, 15277, 15278, 15279, 15280, 15284, 15285, 15304, 16003, 16087, 16190, 17099, 17345, 17432 and 17437 through 18122 may be affected by the problem. Serial numbers 17525, 17697, 17899, 17904, 17905 and 17913 have been corrected.
- 510(k) numbers
- ["K915183","K953904"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28