Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo Perfusion System 8000 Roller Pump, 115V; Model 8000; Catalog No. 16402.

Recall number
Z-1002-2007
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Perfusion System 8000 Roller Pump, 115V; Model 8000; Catalog No. 16402.
Status
Terminated
Initiated
2004-07-01
Posted
2007-07-07
Terminated
2007-12-12
Reason
Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being non-functional.
Root cause
Other
Action
As the complaints were received (between 07/04 and 01/06), the firm visited the customer filing the complaint and repaired the pumps demonstrating the problem. The firm will correct other units as complaints are received. The firm modified its recall strategy and issued a Device Correction Letter dated 6/22/07 notifying all users of the problem
Quantity
728 pumps of all models
Distribution
Worldwide: USA and Bangladesh, Belgium, Bolivia, Canada, Chile, Colombia, Costa Rica, Egypt, Equador, Hong Kong, India, Indonesia, Japan, Latin America, Malaysia, Nicaragua, Philippines, Saudi Arabia, Taipe, Taiwan, United Arab Emirates, Uruguay, and Viet Nam.
Lot, serial or model codes
Serial numbers 10031, 10033, 10137, 10138, 10614, 10663, 10730, 11131, 11281, 11283, 11361, 11548, 11630, 11751, 12110, 12604, 12686, 12938, 13129, 13719, 13826, 13830, 14712, 15090, 15277, 15278, 15279, 15280, 15284, 15285, 15304, 16003, 16087, 16190, 17099, 17345, 17432 and 17437 through 18122 may be affected by the problem. Serial numbers 17525, 17697, 17899, 17904, 17905 and 17913 have been corrected.
510(k) numbers
["K915183","K953904"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register