Roche Molecular Systems, Inc.: cobas 4800 KRAS AMP/DET 24T CE-IVD Mutation Test; CE-IVD 5852170190. Intended for the identification of mutations in codons 12,13 and 61 of the KRAS Gene in DNA derived from formalin-fixed paraffin-embedded human colorectal and non-small cell lung cancer tissues.
- Recall number
- Z-0999-2015
- Recalling firm
- Roche Molecular Systems, Inc.
- Product
- cobas 4800 KRAS AMP/DET 24T CE-IVD Mutation Test; CE-IVD 5852170190. Intended for the identification of mutations in codons 12,13 and 61 of the KRAS Gene in DNA derived from formalin-fixed paraffin-embedded human colorectal and non-small cell lung cancer tissues.
- Status
- Terminated
- Initiated
- 2014-12-18
- Posted
- 2015-06-26
- Terminated
- 2017-02-08
- Reason
- cobas KRAS Mutation Kit T10786 is generating invalid results for the Mutant Control and/or KRAS Calibrator.
- Root cause
- Device Design
- Action
- RMS issued an Urgent Field Safety Notice on 12/18/2014. A Quality Notification was provided to all users of the of cobas KRAS Mutation Kit CE-IVD, lots T13091, T15435, T15475, T14777, and W00389, informing them of the issue and instructing them to discontinue use and discard any remaining inventory immediately.
- Quantity
- 660 pieces
- Distribution
- Foreign Distributed.
- Lot, serial or model codes
- Lots T10786 T13091, T15435, T15475, T14777, and W00389
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OWD
- Firm FEI
- 2243471
- Firm city
- S Branchburg
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28