Skip to the content

FDA Medical Device Recalls (openFDA)

Ossur Americas: Ossur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ Skull Pin 3in TI Sterl 1PK; 520TS/ Skull Pin 2-1/4in TI Sterl 5PK; 520V2S/ Skull Pin 3in TI Sterl 5 PK

Recall number
Z-0997-2019
Recalling firm
Ossur Americas
Product
Ossur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ Skull Pin 3in TI Sterl 1PK; 520TS/ Skull Pin 2-1/4in TI Sterl 5PK; 520V2S/ Skull Pin 3in TI Sterl 5 PK
Status
Terminated
Initiated
2018-06-21
Posted
not on file
Terminated
2020-09-04
Reason
Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
Root cause
Labeling design
Action
On 06/21/2018, Urgent Field Safety Notices were mailed to customers via UPS Next-Day Air requiring a signature of acceptance from the customer. Customers were advised the following: 1. Please pass this notice to those who need to be aware within your organization or to any department areas where the potentially affected devices will be used or transferred. 2. Maintain awareness on this notice for an appropriate period. 3. If you would like to have the product replaced please contact your customer service or sales representative and new product will be provided. Customers can exchange the affected product, if they desire to do so. Customers were asked to distribute the notices to those who need to be aware within their organization or to any department areas where the potentially affected devices were used or transferred. On 01/14/2019, Urgent Medical Device Recall letters were sent via express mail requiring a signature of acceptance from the customer. Customers were advised: Please examine your inventory, quarantine products subject to the recall and contact customer service for a return authorization to return the product. In addition: 1. Please confirm receipt of this recall notice by calling customer service (800-233-6263) or visit https://safetyalert.ossur.com 2. Please pass this notice to those who need to be aware within your organization and to any department areas where the potentially affected devices will be used or transferred (e.g. MR Imaging, Surgical, Emergency, Material Management, Patient) 3. Maintain awareness on this notice for an appropriate period. 4. If you have further distributed this product, please identify your customers and notify them at once of this product alert. We recommend that you include a copy of this product recall. 5. If any of your patients is currently wearing the product, necessary arrangements should be made to ensure the patient is not placed in MR Imaging.
Quantity
372
Distribution
U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
Lot, serial or model codes
All lots of the products manufactured from 06/30/2009 - 12/13/2018  Titanium Pin Label Mistakenly lists Ceramic Tip, Part/Lot numbers: 516TS/ AL18021418 and AL17022109; 520TS/ AL18021417 and AL15091012
510(k) numbers
["K822780"]
PMA numbers
not on file
Product code
JEC
Firm FEI
1000519909
Firm city
Foothill Ranch
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register