Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc. Cardiac Rhythm Disease Management: Medtronic Select 3D Arterial Cannulae. Vented, model 78624 (24 Fr.). Non-Vented Cannulae, model 78724 (24 Fr.). Sterilized using Ethylene Oxide. Non-pyrogenic. The Cannula consists of a curved, angled or beveled tip with or without flange that is permanently attached to a molded, clear, flexible PVC tapered body or wirebound PVC body. The cannula body features a tip orientation line to indicate direction of the cannula tip during cannulation. The proximal end of the cannula includes a connection site or a molded 0.95 cm (3/8 in) barbed connector with or without a molded plastic vent plug or peel cap. The vent plug or peel cap allows air to be vented from the cannula before connection to the perfusion line. These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass.

Recall number
Z-0997-2014
Recalling firm
Medtronic Inc. Cardiac Rhythm Disease Management
Product
Medtronic Select 3D Arterial Cannulae. Vented, model 78624 (24 Fr.). Non-Vented Cannulae, model 78724 (24 Fr.). Sterilized using Ethylene Oxide. Non-pyrogenic. The Cannula consists of a curved, angled or beveled tip with or without flange that is permanently attached to a molded, clear, flexible PVC tapered body or wirebound PVC body. The cannula body features a tip orientation line to indicate direction of the cannula tip during cannulation. The proximal end of the cannula includes a connection site or a molded 0.95 cm (3/8 in) barbed connector with or without a molded plastic vent plug or peel cap. The vent plug or peel cap allows air to be vented from the cannula before connection to the perfusion line. These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass.
Status
Terminated
Initiated
2014-01-30
Posted
2014-02-20
Terminated
2014-10-06
Reason
Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar.
Root cause
Under Investigation by firm
Action
The firm, Medtronic, sent an "Urgent Medical Device Recall" letter dated January, 2014 via UPS on January 30, 2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately identify and quarantine all listed product in your inventory; return all listed product to Medtronic and contact Customer Service at 800 854-3570 and reference RADAR #167826 to initiate a return and credit of product; complete and return the attached Customer Confirmation Certificate via fax to Medtronic at 651-367-0612 to the attention of Customer Focused Quality, and share this notification with others in your organization as appropriate. Select 3D and Select CAP Arterial Cannula are not available for purchase at this time. Please contact your Medtronic sales representative or Medtronic LifeLine Technical Services at 1 877-526-7890 or (+1 763) 526-7890 to assist with alternative cannula selection and any additional questions you may have related to this notification.
Quantity
19483 total (7669 US, 11814 OUS)
Distribution
Worldwide distribution: US (Nationwide) including Puerto Rico, and Internationally to: Algeria, Argentina, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Cyprus, El Salvador, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Jordan, Kuwait, Netherlands, Oman, Paraguay, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, Venezuela, and Yemen.
Lot, serial or model codes
Model 78624, Lot numbers: 2011010432, 2011010514, 2011010515, 2011011141,  2011011543, 2011021907, 2011040344, 2011040946, 2011060613, 2011060865,  2011070535, 2011071749, 2011081635, 2011110330, 2011120304, 2011120305,  2011121133, 2012060424, 2012060431, 2012070886, 2012080324, 2012081353,  2012081928, 2012090013, 2012120521, 2012121021, 2012121414, 2013031142,  2013050555, 2013050805, 2013051225, 2013051640, 2013060249, 2013090499,  2013090570, 2013091332, 2013091924, 201106C140, 201203C844, 201310C579.   Model 78724, Lot numbers: 2011032145, 2011050718, 2011060852, 2011092067,  2011102225, 2012032345, 2012071614, 2012081704, 2012081973, 2013031364,  2013051818, 2013060581, 2013070831, 2013080858, 2013100654, 2013110548,  2014010275, 201108C228, 201203C842.
510(k) numbers
["K000776"]
PMA numbers
not on file
Product code
DWF
Firm FEI
2182208
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register