Covidien Limited: Shiley PED3.0 Cuffless Pediatric Tracheostomy Tube, REF 3.0 PED, Sterile. Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Assembled in Mexico. The product is provided as one each shipped in an inner carton to end customers. The intended use of the device is to bypass upper airway obstructions, provide long term ventilation, and support and/or manage tracheal/bronchial secretions.
- Recall number
- Z-0997-2009
- Recalling firm
- Covidien Limited
- Product
- Shiley PED3.0 Cuffless Pediatric Tracheostomy Tube, REF 3.0 PED, Sterile. Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Assembled in Mexico. The product is provided as one each shipped in an inner carton to end customers. The intended use of the device is to bypass upper airway obstructions, provide long term ventilation, and support and/or manage tracheal/bronchial secretions.
- Status
- Terminated
- Initiated
- 2009-01-14
- Posted
- 2009-03-04
- Terminated
- 2012-05-21
- Reason
- Difficulty inserting an obturator and/or suction catheter into the tracheostomy tube.
- Root cause
- Finished device change control
- Action
- Consignees were sent an "Urgent Device Alert" letter dated 01/14/2009, by mail via Federal Express overnight delivery and instructed to evaluate any of the affected lot numbers of tracheostomy tubes prior to placement into a patient's stoma. If either the obturator or the desired sterile suction catheter does not fit properly into the tube, the tube must not be inserted into the patient. Further instructions are available at Nellcor Technical Services by calling 1-800-635-5267, option 3, then option 1. New product was expected by mid February, 2009, to address the issue.
- Quantity
- 8,931 units
- Distribution
- Nationwide Distribution, including Puerto Rico.
- Lot, serial or model codes
- Lot Numbers: 0806000266, 0806000858, 0806000859, 0806001328, 0807000555, 0807001444, 0807002502, 0808000387, 0808000431, 0808002285, 0810000288, 0810000289, 0810000864, 0810000865, 0810001201, 0811000228, 0811000522, 0811000523, 0811000546, 0811000547 and 0811001875.
- 510(k) numbers
- ["K945513"]
- PMA numbers
- not on file
- Product code
- JOH
- Firm FEI
- 2936999
- Firm city
- Boulder
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28