Siemens Medical Solutions USA, Inc: syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.

Recall number
Z-0995-2022
Recalling firm
Siemens Medical Solutions USA, Inc
Product
syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
Status
Terminated
Initiated
2022-03-07
Posted
not on file
Terminated
2023-12-19
Reason
Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
Root cause
Software change control
Action
4/4/22 update: Urgent Medical Device Correction letter has not been issued. The proposed letter informs customers of software issue and workaround to resolve issue. It also requires electronic signature to acknowledge receipt of the notification. If you have additional questions contact the service organization at 1-800-888-7436.
Quantity
1 (US)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Switzerland, United Kingdom.
Lot, serial or model codes
Partial UDI: 04056869051314, Serial Number 112261
510(k) numbers
["K211373"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register