FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
- Recall number
- Z-0995-2022
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
- Status
- Terminated
- Initiated
- 2022-03-07
- Posted
- not on file
- Terminated
- 2023-12-19
- Reason
- Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
- Root cause
- Software change control
- Action
- 4/4/22 update: Urgent Medical Device Correction letter has not been issued. The proposed letter informs customers of software issue and workaround to resolve issue. It also requires electronic signature to acknowledge receipt of the notification. If you have additional questions contact the service organization at 1-800-888-7436.
- Quantity
- 1 (US)
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Switzerland, United Kingdom.
- Lot, serial or model codes
- Partial UDI: 04056869051314, Serial Number 112261
- 510(k) numbers
- ["K211373"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06