Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: AMSCO 3000 Series Washer/Disinfector Model # 3052

Recall number
Z-0995-2019
Recalling firm
Steris Corporation
Product
AMSCO 3000 Series Washer/Disinfector Model # 3052
Status
Terminated
Initiated
2018-12-20
Posted
not on file
Terminated
2019-07-22
Reason
The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
Root cause
Software Manufacturing/Software Deployment
Action
Urgent Device Field Correction notification letters dated 12/20/18 were sent to customers.
Quantity
11
Distribution
The products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.
Lot, serial or model codes
Serial Numbers 3602014006 3631618006 3628818009 3629518008 3633018010 3626718010 3630918008 3630918010 3617717013 3620016019 3623516006
510(k) numbers
not on file
PMA numbers
not on file
Product code
MEC
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register