Skip to the content

FDA Medical Device Recalls (openFDA)

Kimberly-Clark Corporation: KIMGUARD Container Filters, 7.5in (19cm), Round and 9in. X 9in. (23cm x 23cm), Square. Intended to be used to enclose another medical device that is to be sterilized by a health care provider.

Recall number
Z-0995-2013
Recalling firm
Kimberly-Clark Corporation
Product
KIMGUARD Container Filters, 7.5in (19cm), Round and 9in. X 9in. (23cm x 23cm), Square. Intended to be used to enclose another medical device that is to be sterilized by a health care provider.
Status
Terminated
Initiated
2013-03-06
Posted
2013-03-26
Terminated
2013-09-16
Reason
Some individual container filter units might contain thin areas that may potentially compromise the ability of the filter to maintain a sterile barrier.
Root cause
Component design/selection
Action
All US distributors and clinics were sent an Urgent: Voluntary Product Recall letter via Federal Express, and the Canadian importer/distributor was also notified with the same letter via Federal Express and e-mail. Distributors were instructed to contact their customers by sending a copy of the letter. The letter identified the affected product and stated the reason for the recall. The recall notification includes a Response Sheet that allows the customer (end-user) to indicate the quantity of affected product they have in their inventory or indicate whether they do not have any affected product. The customer is instructed to immediately cease further distribution of the product, physically quarantine and destroy the product, and fax the completed Response Sheet to the Kimberly-Clark Recall Coordinator.
Quantity
Code 68507 - 6420 cases; Code 68509 - 1281 cases
Distribution
Worldwide Distribution -- USA, including the states of AL, AK, AZ, AR, CA, CO, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NV, NJ, NM, NY, NC, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, Belgium, Hong Kong, Japan, Portugal, and Singapore.
Lot, serial or model codes
Product Code 68507 (Round) Case Lot Numbers MF1284XXX, MF1287XXX, MF1294XXX, MF1316XXX, MF2012XXX, MF2055XXX, MF2086XXX, MF2098XXX, MF2115XXX, MF2146XXX, MF2178XXX, MF2188XXX, MF2196XXX, MF2210XXX, MF2228XXX, MF2243XXX, MF2264XXX, MF2271XXX, MF2285XXX, MF2328XXX, and MF2338XXX.   Product Code 68509 (Square) Case Lot Numbers MF1279XXX, MF1287XXX, and MF1288XXX.
510(k) numbers
["K881471"]
PMA numbers
not on file
Product code
FRG
Firm FEI
1000015879
Firm city
Roswell
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register