FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio Max is a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio Max system is not meant for mammography. The Ysio Max uses integrated or portable digital detectors for generating diagnostic images by converting x-rays into electronic signals. Ysio Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
- Recall number
- Z-0994-2015
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio Max is a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio Max system is not meant for mammography. The Ysio Max uses integrated or portable digital detectors for generating diagnostic images by converting x-rays into electronic signals. Ysio Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
- Status
- Terminated
- Initiated
- 2014-12-15
- Posted
- 2015-01-16
- Terminated
- 2017-01-26
- Reason
- Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.
- Root cause
- Software design
- Action
- The firm, Siemens, sent a "Safety Advisory Notice - XP080/14/S" letter dated December 8, 2014, to end users that identified the product, problem, and actions to be taken. Siemens provided instructions to avoid the issue and indicated a software fix would be released as a permanent resolution. The customers were instructed observe this safety notice and comply with the corresponding measure until the update has been fully completed. If you have any questions, contact the Regulatory Technical Specialist at 610-219-2119.
- Quantity
- 11
- Distribution
- US Distribution to states of: ID, IL, MI, MN, MO, ND and PA.
- Lot, serial or model codes
- Material 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 serial numbers24077 24051 24038 24059 24066 24015 24026 24058 24039 24090 24075
- 510(k) numbers
- ["K062623","K111292","K133259"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28