FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S Model Number: TB-0535FCS Catalog Number/Product Code: R5000687 Software Version: N/A Product Description: Thunderbeat instrument is a sterile, single use item which is intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (WSG-400), and the Thunderbeat Transducer, (TD-TB400).
- Recall number
- Z-0993-2025
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S Model Number: TB-0535FCS Catalog Number/Product Code: R5000687 Software Version: N/A Product Description: Thunderbeat instrument is a sterile, single use item which is intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (WSG-400), and the Thunderbeat Transducer, (TD-TB400).
- Status
- Open, Classified
- Initiated
- 2024-11-25
- Posted
- 2025-01-21
- Terminated
- not on file
- Reason
- The probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
- Root cause
- Under Investigation by firm
- Action
- On December 20, 2024, URGENT FIELD SAFETY NOTICE: ADVISORY letters were sent to customers. Actions Required: Olympus requires you to take the following actions: 1. Carefully read the content of this FSN. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this FSN and the relevant IFU. 3. Olympus requests that you acknowledge receipt of this FSN through our web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number "0462 c. Complete the form as instructed and include your account ID number: [XXXXXXXX] d. Indicating in the comments if you would like to receive in-service training to support the correct use of this device. 4. If you have further distributed this product, identify your customers, and forward this FSN to them. Olympus requests that you report any complaints, including probe damage or pad detachment, to Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1), and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@olympus.com.
- Quantity
- 495,865 units (96,253 US, 399,612 OUS)
- Distribution
- US Nationwide distribution and International: Algeria, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Belize, Benin, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Nepale, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, Utd.Arab Emir., Vietnam, Yemen.
- Lot, serial or model codes
- Model Number: TB-0535FCS; Catalog Number/Product Code: R5000687; UDI: 04953170409677 and 04953170383540; Lots: All
- 510(k) numbers
- ["K211838"]
- PMA numbers
- not on file
- Product code
- LFL
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28