Arstasis, Inc: Arstasls One Access System; Models: AAD100, AAD101, Part Numbers: FG-02279, FG-03010; The system consists of the following components: (1) Device Sheath/Anchor, (2) Shaft and (3) Handle with control features. Product is manufactured and distributed by Arstasis Inc., San Carlos, CA The Arstasis One Access System is a sterile, single use system intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The System is Indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using SF or 6F Introducer sheaths.

Recall number
Z-0993-2011
Recalling firm
Arstasis, Inc
Product
Arstasls One Access System; Models: AAD100, AAD101, Part Numbers: FG-02279, FG-03010; The system consists of the following components: (1) Device Sheath/Anchor, (2) Shaft and (3) Handle with control features. Product is manufactured and distributed by Arstasis Inc., San Carlos, CA The Arstasis One Access System is a sterile, single use system intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The System is Indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using SF or 6F Introducer sheaths.
Status
Terminated
Initiated
2010-10-19
Posted
2011-02-04
Terminated
2011-02-16
Reason
Device has the potential to fracture and sheath separation during use.
Root cause
Device Design
Action
The firm, Arstasis, initiated a market withdrawal and sent notification via letters dated October 19, 2010, to customers informing the customers that Arstasis is implementing a product improvement and will replace all existing inventory that the customer has on their shelf with the improved product. The customers were instructed to work closely with their local Arstasis territory manager to ensure that their product replacement is handled satisfactorily, and complete and return the PRODUCT REPLACEMENT and PRODUCT RETURN forms to Arstasis Customer Service via fax at 1-877-688-8417 or email to customerservice@arstasis.com. If you have any questions or concerns, contact Customer Service toll free at 1-877-594-4545.
Quantity
330 units
Distribution
Worldwide Distribution
Lot, serial or model codes
Lot Numbers: 09I10268, 1OC26337, 09J06281, 10C12334
510(k) numbers
["K091006","K100615","K102728"]
PMA numbers
not on file
Product code
DYB
Firm FEI
3008493244
Firm city
Redwood City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register