FDA Medical Device Recalls (openFDA)
Randox Laboratories Ltd.: RX Series Copper (Cu) Assay Ref. Number CU2340
- Recall number
- Z-0992-2024
- Recalling firm
- Randox Laboratories Ltd.
- Product
- RX Series Copper (Cu) Assay Ref. Number CU2340
- Status
- Open, Classified
- Initiated
- 2023-11-29
- Posted
- 2024-02-06
- Terminated
- not on file
- Reason
- Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.
- Root cause
- Component design/selection
- Action
- RANDOX Distributors issued Medical Device Correction Letter (REC704) on December 12, 2023 to US consignees. Letter states reason for recall, health risk and action to take: Review your instrument testing order in line with the Carryover Avoidance Technical Bulletin (RXTB-0148) " Update the RX user manual with the updated Carryover Avoidance Technical Bulletin (RXTB-0148) and ensure all operators are aware of the recommendations. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
- Quantity
- 3,053 kits (2 US, 3,051 OUS)
- Distribution
- US Nationwide distribution in the states of SC and OK.
- Lot, serial or model codes
- Catalogue Number/GTIN: CU2340 05055273201949; Lot/batch: 592278, 610661, 610669, 629021, 647201, 657944;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JKZ
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as