Ortho-Clinical Diagnostics: VITROS Chemistry Products Na+ Slides, Catalog Number 8379034 For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Recall number
Z-0992-2019
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Products Na+ Slides, Catalog Number 8379034 For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Status
Terminated
Initiated
2019-01-22
Posted
not on file
Terminated
2022-06-10
Reason
Imprecision in the affected lot may produce negatively or positively biased patient sample and quality control fluid results.
Root cause
Under Investigation by firm
Action
The firm, Ortho Clinical Diagnostics, sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letters dated 1/22/19 to is customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " If you have an alternate lot of VITROS Na+ Slides, discontinue using and discard all remaining inventory of Lot 4221-0998-3223. " If you do NOT have an alternate lot, consult with your medical or laboratory director to determine if it is acceptable to continue using the affected lot until your replacement order arrives. " Until you receive your replacement order: o Performing Osmolality calculations on your VITROS System is not recommended. o If you continue using the affected lot, it is strongly recommended to take additional action to detect imprecision, such as performing duplicate testing for each sample. Repeat testing if there is a discrepancy between the results obtained in duplicate testing (i.e., +/- 1.4 mmol/L). o Discard all remaining inventory when your replacement order arrives. An adequate supply of replacement product is available and Ortho will expedite all replacement orders. " Please complete the enclosed Confirmation of Receipt form by e-Mail to RA-OCDUS-CONFIRMAD@ORTHOCLINICALDIAGNOSTICS.COM or Fax to: 1.888.557.3759 or 1.585.453.4110. Ortho will credit your account or send a replacement order for the discarded slides. " Forward the information in this notification if you have distributed this product outside of your facility. If you have questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Quantity
1448
Distribution
US Distribution to states of: AL, CA, CT, FL, HI, IA, ID, IL, LA, MI, MO, MS, NC, NM, NY, OK, PA, TX, WV, and WY; and internationally to: Canada.
Lot, serial or model codes
Lot 4221-0998-3223
510(k) numbers
["K081411"]
PMA numbers
not on file
Product code
JGS
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register