Biocardia, Inc.: MORPH ACCESSPRO Steerable Introducer; 6F introducer, 90 cm working length: Sterile in unopened, undamaged pouch. Sterilized by EtO gas, Non-pyrogenic: BioCardia Inc. San Carlos, CA 94070 Intended to provide a pathway through which medical instruments such as balloon dilation catheters, guidewires and other therapeutic devices may be introduced into the peripheral vasculature.

Recall number
Z-0991-2014
Recalling firm
Biocardia, Inc.
Product
MORPH ACCESSPRO Steerable Introducer; 6F introducer, 90 cm working length: Sterile in unopened, undamaged pouch. Sterilized by EtO gas, Non-pyrogenic: BioCardia Inc. San Carlos, CA 94070 Intended to provide a pathway through which medical instruments such as balloon dilation catheters, guidewires and other therapeutic devices may be introduced into the peripheral vasculature.
Status
Terminated
Initiated
2014-02-07
Posted
2014-02-19
Terminated
2014-04-03
Reason
Tyvek packaging may be worn, causing loss of package integrity.
Root cause
Process control
Action
Sales Representative called each affected hospital on February 7, 2014, to notify them of the recall. BioCardia has initiated a recall of Morph AccessPro due to a potential compromise of the sterile barrier. Use of this product may result in patient infection. Sales representatives contacted hospital sites identified in Attachment 3 by telephone. Script provided as follows: "BioCardia has initiated a recall of Morph Access Pro due to a potential compromise of the sterile barrier. Use of this product may result in patient infection. Our records indicate that we sold you. The recall is limited to these units in this lot sold to you. We would like to confirm if you still have the unit(s). If you do, we ask that you quarantine the unit(s) and our customer service department will contact you to coordinate the return of the unit(s)." For further questions please call (650) 226-0133.
Quantity
11 units
Distribution
US Distribution including the states of FL, MA, WY. OH, MI and KS.
Lot, serial or model codes
Catalog number MAP690; Lot numbers 01162 and 01186.
510(k) numbers
["K090999"]
PMA numbers
not on file
Product code
DYB
Firm FEI
3003610571
Firm city
San Carlos
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register