FDA Medical Device Recalls (openFDA)

Southmedic, Inc.: Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0615, Size #15, General Surgical Use, Sterile, unless package is damaged or opened., Southmedic Inc., 50 Alliance Blvd., Barrie, ON., Canada L4M 5K3 www.southmedic.com, Made in Canada.

Recall number
Z-0990-2011
Recalling firm
Southmedic, Inc.
Product
Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0615, Size #15, General Surgical Use, Sterile, unless package is damaged or opened., Southmedic Inc., 50 Alliance Blvd., Barrie, ON., Canada L4M 5K3 www.southmedic.com, Made in Canada.
Status
Terminated
Initiated
2010-09-03
Posted
2011-01-31
Terminated
2011-05-18
Reason
Scalpels labeled as sterile have the potential to be unsealed and therefore not sterile.
Root cause
Packaging process control
Action
On 9/3/2010, the firm notified consignees about the recall by facsimile letter. The letter instructed consignees to return scalpels for credit or replacement. The firm enlarged the scope of the recall on 11/15/2010. An additional 927 cases were recalled. This includes 11 additional lots. The firm sent out URGENT MEDICAL DEVICE RECALL letters dated 11/15/2010 to its direct accounts. The letter identified that affected product and the reason for the recall. The consignees were instructed to check their inventory, complete and return the attached Voluntary Recall Notification Confirmation Form, return the affected product to the manufacturer. Product return question should be directed towards Marc Pauze at 800-463-7146 ext 299 and regulatory questions should be directed towards Tish Anger at 800-463-7146 ext 342.
Quantity
1330 cases all products
Distribution
Worldwide Distribution -- USA, including states of TX, MO, NJ, LA, IL, NY, KY, PA, VA, WA, CA, MA, MD, IN, NV, TN, NC, AZ, IA, WI, UT, NE, GA, FL, MN, ME, OH, SD, and KS and countries of Australia and Canada.
Lot, serial or model codes
Lot numbers: 070910, 081610, 90110, 91410, and 93010.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GDX
Firm FEI
3001717360
Firm city
Barrie Canada
Firm state
Ontario
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register