Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical MD, Inc.: Deltec Cozmo Insulin Pump, Model 1700 containing 1800 software. Deltec Cozmo Insulin Pumps are ambulatory electromechanical pumps used for delivering insulin to patients with diabetes. The pump is used mainly by homecare patients, but can also be used in a healthcare facility.

Recall number
Z-0989-2009
Recalling firm
Smiths Medical MD, Inc.
Product
Deltec Cozmo Insulin Pump, Model 1700 containing 1800 software. Deltec Cozmo Insulin Pumps are ambulatory electromechanical pumps used for delivering insulin to patients with diabetes. The pump is used mainly by homecare patients, but can also be used in a healthcare facility.
Status
Terminated
Initiated
2008-11-20
Posted
2009-04-16
Terminated
2012-06-05
Reason
Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered was not accurately displayed by the Pump. The amount of extended Bolus displayed on the Pump Home Screen 2 and in the Bolus Summary Report is less than what w
Root cause
Software design
Action
Consignees were sent a Smiths Medical "Urgent Device Recall" letter dated November 20, 2008. The letter was addressed to Distributors, Clinicians, Patients and other Customers. The letter included a Description of the Problem, Advice on Action to be Taken by the User and requested the return of the Confirmation Form. In particular, users were given the option to check the "Replacement Pump" option on the Confirmation Form initiating the process for pump replacement with a recertified pump followed by a return of the defective product. The Confirmation Form can be faxed to 1-800-628-6322 in the U.S. or 1-651-628-7485 outside the U.S., completed at www.cozmore.com/notice or calling 1-800-501-5748 in the U.S.
Quantity
586
Distribution
Worldwide Distribution --- Including states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MI, MA, MD, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and DC and countries of AUSTRALIA, NEW ZEALAND, UNITED KINGDOM, IRELAND, CANADA, CHINA, GREECE, POLAND, SLOVENIA, SWEDEN, ISRAEL and BRAZIL.
Lot, serial or model codes
Product Numbers: 21-1705-79 21-1705-89 21-1706-79 21-1706-89 21-1707-79 21-1707-89 21-1715-19 21-1716-19 and 21-1717-19.
510(k) numbers
["K062323"]
PMA numbers
not on file
Product code
LZG
Firm FEI
2183502
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register