Getinge Usa Sales Inc: Flow-i C30 Anesthesia System, model 6677300

Recall number
Z-0988-2024
Recalling firm
Getinge Usa Sales Inc
Product
Flow-i C30 Anesthesia System, model 6677300
Status
Open, Classified
Initiated
2023-12-11
Posted
2024-02-06
Terminated
not on file
Reason
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Root cause
Software design
Action
An Urgent Medical Device Correction notice dated December 21, 2023 was issued via FedEx. The letter advises customers to examine inventory immediately to determine if you have any affected product. Directions are outlined for 1) Users with system version 4.8 and above with a Flow Anesthesia System connected to Connected Services directly via the Ethernet port on the Flow Anesthesia System (i.e., without a Getinge Connect module) and for 2) Users with system version 4.7 and below with Flow Anesthesia System connected to Remote Services via ethernet cable. Customers are asked to complete and return the MEDICAL DEVICE - CORRECTION RESPONSE FORM. Information should be forwarded to all current and potential system users with in your hospital / facility. Distributors are to forward the notification to customers for appropriate action. Contact your Getinge representative or call Getinge Customer Service at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) with questions.
Quantity
4 (US)
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
Lot, serial or model codes
UDI-DI: 07325710001363, serial numbers (US): 1058, 1060, 1167, 1201.
510(k) numbers
["K133958"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
3013876692
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register