Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Hearing Instruments, Inc: Siemens Prisma 2K Hearing Aid Product Usage: A behind the ear (BTE) hearing aid developed for pediatric patients.

Recall number
Z-0988-2014
Recalling firm
Siemens Hearing Instruments, Inc
Product
Siemens Prisma 2K Hearing Aid Product Usage: A behind the ear (BTE) hearing aid developed for pediatric patients.
Status
Terminated
Initiated
2005-07-01
Posted
2014-02-18
Terminated
2017-07-06
Reason
Siemens became aware of an issue where an infant or child could open the tamper resistant battery compartment of the Siemens Prisma 2K hearing aid under certain unintended use conditions.
Root cause
Device Design
Action
Siemens sent a notice letter to affected customers in 7/2005 and 12/2005. The letter identified the affected product, problem and actions to be taken. Siemens will install an additional security measure to affected product at no charge. For questions contact your Sales Representative at 800-766-4500.
Quantity
738 units
Distribution
USA Nationwide Distribution
Lot, serial or model codes
Part numbers 1) 7773182 2) 7244028 3) 8757374 4) 8747086 5) 10056937 6) 10056938
510(k) numbers
not on file
PMA numbers
not on file
Product code
LEZ
Firm FEI
2217809
Firm city
Piscataway
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register