Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic product for the calibration of the Alkaline Phosphatase method on the Dimension Vista" System.

Recall number
Z-0988-2013
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic product for the calibration of the Alkaline Phosphatase method on the Dimension Vista" System.
Status
Terminated
Initiated
2012-10-10
Posted
2013-03-25
Terminated
2014-01-08
Reason
Correlation slope bias and accuracy shift
Root cause
Nonconforming Material/Component
Action
Siemens sent a Urgent Medical Device Correction letter dated October 16, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to calibrate with newer lots of Dimension Vista(R) ALP Calibrator as they become available, beginning with lot 2GD011. Customers were also asked to discontinue use and discard any remaining inventory of lot 2AD052. Customers further instructed to complete the attached response form and fax to (302) 631-8467, to indicate they had received the information and to facilitate their no charge replacement. Further questions please call (302) 631-6311.
Quantity
1985 cartons
Distribution
Nationwide distribution: USA including Puerto Rico and the states of AL, OH, MO, GA, IL, CT, IA, SC, VA, TX, FL, NY, NC, CA, NM, AZ, CA, MI, NJ, PR, MS, NE, WV, PA, MT, MA, MD, KS, NH, TN, VT, LA, CO, ID, WA, NH, DE and OR.
Lot, serial or model codes
Lot 2AD052 Catolog KC330
510(k) numbers
["K061818"]
PMA numbers
not on file
Product code
JIT
Firm FEI
2517506
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register