FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc.: Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic product for the calibration of the Alkaline Phosphatase method on the Dimension Vista" System.
- Recall number
- Z-0988-2013
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic product for the calibration of the Alkaline Phosphatase method on the Dimension Vista" System.
- Status
- Terminated
- Initiated
- 2012-10-10
- Posted
- 2013-03-25
- Terminated
- 2014-01-08
- Reason
- Correlation slope bias and accuracy shift
- Root cause
- Nonconforming Material/Component
- Action
- Siemens sent a Urgent Medical Device Correction letter dated October 16, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to calibrate with newer lots of Dimension Vista(R) ALP Calibrator as they become available, beginning with lot 2GD011. Customers were also asked to discontinue use and discard any remaining inventory of lot 2AD052. Customers further instructed to complete the attached response form and fax to (302) 631-8467, to indicate they had received the information and to facilitate their no charge replacement. Further questions please call (302) 631-6311.
- Quantity
- 1985 cartons
- Distribution
- Nationwide distribution: USA including Puerto Rico and the states of AL, OH, MO, GA, IL, CT, IA, SC, VA, TX, FL, NY, NC, CA, NM, AZ, CA, MI, NJ, PR, MS, NE, WV, PA, MT, MA, MD, KS, NH, TN, VT, LA, CO, ID, WA, NH, DE and OR.
- Lot, serial or model codes
- Lot 2AD052 Catolog KC330
- 510(k) numbers
- ["K061818"]
- PMA numbers
- not on file
- Product code
- JIT
- Firm FEI
- 2517506
- Firm city
- Tarrytown
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28