Philips Medical Systems (Cleveland) Inc: Gamma camera systems: AXIS SPECT Systems & IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips Healthcare. Used to do nuclear medicine diagnostic imaging of various body organs.

Recall number
Z-0988-2012
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Gamma camera systems: AXIS SPECT Systems & IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips Healthcare. Used to do nuclear medicine diagnostic imaging of various body organs.
Status
Terminated
Initiated
2011-11-18
Posted
2012-02-10
Terminated
2013-02-22
Reason
Philips was informed that the detector buckets may experience weakness over time. As a result, this could cause the detector bucket to separate from the mounting plate of the gantry and eventually result in the detector falling.
Root cause
Nonconforming Material/Component
Action
On 11/15/2011 Philips sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected product, the problem, the hazard involved, and the actions to be taken by users and the Philips. Customers are to place the letter in their User Documentation. Philips service will arrange a time to inspect customers' systems and conduct the appropriate field safety correction. Customers are to contact their local Philips representative if they need further support or information. In the US and Canada, customers can contact Philips Healthcare Customer Care Solutions Center at 1-800-722-9377.
Quantity
795 units
Distribution
Worldwide Distribution -- USA, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PRO, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY and the territory of Puerto Rico; and countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Egypt, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, & United Kingdom.
Lot, serial or model codes
System Code #, 882130, Serial #, 10, 97, 101, 103, 104, 107-112, 116, 117, 119-121, 123, 125, 127, 130, 132, 133, 135, 138, 142-144, 149-151, 153-158, 162-164, 166, 169-173, 180, 181, 185, 186, 189, 193-204, 214, 215, 218, 219, 221-223, 227, 229-231, 233-237, 240, 242, 244-246, 249-251, 260-262, 264, 267, 272, 273, 275, 277, 280, 281, 283, 285, 291-293, 295, 297-300, 305, 308-310, 312, 316, 319-322, 326-332, 334-336, 338-340, 346, 347, 349, 351, 352, 354, 355, 358-363, 368-370, 374, 377-379, 385, 387, 388, 390-396, 401-407, 411, 412, 416-419, 426-432, 434, 437, 438, 442, 443, 445-447, 449, 452, 457-462, 464, 465, 468, 471, 474, 480-485, 487, 490-493, 495, 496, 499, 502, 503, 505, 508, 509, 511, 513-518, 520, 521, 523, 529, 530, 534, 543-546, 548-560, 562, 563, 569-572, 574-576, 580, 582, 583, 585, 586, 590, 594, 595, 597, 602-607, 611, 612, 618-621, 624-628, 630, 635, 636, 640, 641, 647-650, 654, 656-658, 660-665, 668-673, 675, 679, 681-684, 686, 687, 690-695, 697-699, 701-708, 711-719, 723-729, 733-736, 738, 739, 743-746, 749-753, 755, 756, 759, 760, 762, 764-767, 771-773, 776-778, 780, 781, 786, 789, 790-795, 797, 799, 801, 803-806, 808, 809, 811-814, 818, 820-822, 825-827, 831-834, 836, 837, 839, 840, 841, 843-845, 848, 849, 851-855, 857, 858, 860-862, 865-869, 872-874, 876-881, 884-886, 889, 893, 894, 900-903, 907-922, 924, 927-930, 932, 934, 938-941, 943, 944, 946-948, 951, 953, 954, 958-963, 965-967, 970, 972-980, 982, 983, 985, 986, 991-994, 996-1000, 1003-1008, 1010-1013, 1016, 1017, 1019, 1022, 1023, 2007, 2008, 211037, 2008-0068, 555A, & 778A. System Code #882140, Serial # 102, 134, 141, 146, 147, 159, 160, 161, 167, 175, 176, 190, 209, 210, 216, 225, 226, 257, 259, 268-270, 274, 289, 290, 302, 304, 314, 315, 318, 323, 333, 341, 343, 344, 348, 350, 364, 366, 381, 386, 409, 414, 420, 421, 423, 436, 441, 455, 463, 469, 472, 473, 488, 489, 497, 498, 501, 531, 532, 536-539, 565, 566, 578, 588, 589, 591, 596, 598-600, 608, 610, 613, 615, 616, 622, 631, 632, 634, 637, 653, 666, 674, 676-678, 685, 689, 700, 709, 710, 731, 732, 741, 742, 748, 757, 763, 768, 774, 785, 802, 810, 816, 819, 824, 828, 829, 830, 838, 846, 856, 863, 864, 870, 871, 875, 884, 891, 892, 898, 904, 905, 931, 933, 936, 942, 945, 949, 950, 952, 955-957, 964, 968, 981, 989, 990, 995, 1001, 1002, 1009, 1015, 1018, & 1021. System Code #882130/40, Serial #184, 730, 122, 136, 254, 255, 284, 614, 696, 874, 906, & 984. System Code #882140/882130, Serial #113, 148, 152, 178, 179, 187, 188, 217, 238, 243, 247, 248, 278, 288, 345, 367, 371, 373, 380, 384, 389, 397, 397, 399, 408, 413, 415, 435, 439, 439, 440, 453, 477-479, 486, 504, 512, 525, 526, 528, 533, 535, 540, 547, 564, 567, 568, 577, 579, 581, 638, 639, 644, 645, 651, 652, 659, 667, 680, 747, 758, 815, 850, 859, 882, 883, 890, 896, 897, 899, 925, & 969.
510(k) numbers
["K964712"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register