Philips Medical Systems (Cleveland) Inc: Gamma camera systems: AXIS SPECT Systems & IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips Healthcare. Used to do nuclear medicine diagnostic imaging of various body organs.
- Recall number
- Z-0988-2012
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Gamma camera systems: AXIS SPECT Systems & IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips Healthcare. Used to do nuclear medicine diagnostic imaging of various body organs.
- Status
- Terminated
- Initiated
- 2011-11-18
- Posted
- 2012-02-10
- Terminated
- 2013-02-22
- Reason
- Philips was informed that the detector buckets may experience weakness over time. As a result, this could cause the detector bucket to separate from the mounting plate of the gantry and eventually result in the detector falling.
- Root cause
- Nonconforming Material/Component
- Action
- On 11/15/2011 Philips sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected product, the problem, the hazard involved, and the actions to be taken by users and the Philips. Customers are to place the letter in their User Documentation. Philips service will arrange a time to inspect customers' systems and conduct the appropriate field safety correction. Customers are to contact their local Philips representative if they need further support or information. In the US and Canada, customers can contact Philips Healthcare Customer Care Solutions Center at 1-800-722-9377.
- Quantity
- 795 units
- Distribution
- Worldwide Distribution -- USA, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PRO, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY and the territory of Puerto Rico; and countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Egypt, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, & United Kingdom.
- Lot, serial or model codes
- System Code #, 882130, Serial #, 10, 97, 101, 103, 104, 107-112, 116, 117, 119-121, 123, 125, 127, 130, 132, 133, 135, 138, 142-144, 149-151, 153-158, 162-164, 166, 169-173, 180, 181, 185, 186, 189, 193-204, 214, 215, 218, 219, 221-223, 227, 229-231, 233-237, 240, 242, 244-246, 249-251, 260-262, 264, 267, 272, 273, 275, 277, 280, 281, 283, 285, 291-293, 295, 297-300, 305, 308-310, 312, 316, 319-322, 326-332, 334-336, 338-340, 346, 347, 349, 351, 352, 354, 355, 358-363, 368-370, 374, 377-379, 385, 387, 388, 390-396, 401-407, 411, 412, 416-419, 426-432, 434, 437, 438, 442, 443, 445-447, 449, 452, 457-462, 464, 465, 468, 471, 474, 480-485, 487, 490-493, 495, 496, 499, 502, 503, 505, 508, 509, 511, 513-518, 520, 521, 523, 529, 530, 534, 543-546, 548-560, 562, 563, 569-572, 574-576, 580, 582, 583, 585, 586, 590, 594, 595, 597, 602-607, 611, 612, 618-621, 624-628, 630, 635, 636, 640, 641, 647-650, 654, 656-658, 660-665, 668-673, 675, 679, 681-684, 686, 687, 690-695, 697-699, 701-708, 711-719, 723-729, 733-736, 738, 739, 743-746, 749-753, 755, 756, 759, 760, 762, 764-767, 771-773, 776-778, 780, 781, 786, 789, 790-795, 797, 799, 801, 803-806, 808, 809, 811-814, 818, 820-822, 825-827, 831-834, 836, 837, 839, 840, 841, 843-845, 848, 849, 851-855, 857, 858, 860-862, 865-869, 872-874, 876-881, 884-886, 889, 893, 894, 900-903, 907-922, 924, 927-930, 932, 934, 938-941, 943, 944, 946-948, 951, 953, 954, 958-963, 965-967, 970, 972-980, 982, 983, 985, 986, 991-994, 996-1000, 1003-1008, 1010-1013, 1016, 1017, 1019, 1022, 1023, 2007, 2008, 211037, 2008-0068, 555A, & 778A. System Code #882140, Serial # 102, 134, 141, 146, 147, 159, 160, 161, 167, 175, 176, 190, 209, 210, 216, 225, 226, 257, 259, 268-270, 274, 289, 290, 302, 304, 314, 315, 318, 323, 333, 341, 343, 344, 348, 350, 364, 366, 381, 386, 409, 414, 420, 421, 423, 436, 441, 455, 463, 469, 472, 473, 488, 489, 497, 498, 501, 531, 532, 536-539, 565, 566, 578, 588, 589, 591, 596, 598-600, 608, 610, 613, 615, 616, 622, 631, 632, 634, 637, 653, 666, 674, 676-678, 685, 689, 700, 709, 710, 731, 732, 741, 742, 748, 757, 763, 768, 774, 785, 802, 810, 816, 819, 824, 828, 829, 830, 838, 846, 856, 863, 864, 870, 871, 875, 884, 891, 892, 898, 904, 905, 931, 933, 936, 942, 945, 949, 950, 952, 955-957, 964, 968, 981, 989, 990, 995, 1001, 1002, 1009, 1015, 1018, & 1021. System Code #882130/40, Serial #184, 730, 122, 136, 254, 255, 284, 614, 696, 874, 906, & 984. System Code #882140/882130, Serial #113, 148, 152, 178, 179, 187, 188, 217, 238, 243, 247, 248, 278, 288, 345, 367, 371, 373, 380, 384, 389, 397, 397, 399, 408, 413, 415, 435, 439, 439, 440, 453, 477-479, 486, 504, 512, 525, 526, 528, 533, 535, 540, 547, 564, 567, 568, 577, 579, 581, 638, 639, 644, 645, 651, 652, 659, 667, 680, 747, 758, 815, 850, 859, 882, 883, 890, 896, 897, 899, 925, & 969.
- 510(k) numbers
- ["K964712"]
- PMA numbers
- not on file
- Product code
- KPS
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28