Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Oncentra External Beam Oncentra Brachy Product Usage: Oncentra is radiation therapy planning software designed to analyze and plan radiation treatment is three dimensions for the purpose of treating patients with cancer.

Recall number
Z-0987-2017
Recalling firm
Elekta, Inc.
Product
Oncentra External Beam Oncentra Brachy Product Usage: Oncentra is radiation therapy planning software designed to analyze and plan radiation treatment is three dimensions for the purpose of treating patients with cancer.
Status
Terminated
Initiated
2016-12-23
Posted
not on file
Terminated
2021-07-16
Reason
Cross profile for Varian 60 degree wedge shows "horns."
Root cause
Software design
Action
Elekta sent an Urgent Important Safety Notification letter dated December 23, 2016 to affected customers. The notice informs users of the specific product and version numbers affected by the issue and any work arounds that can be used to avoid the issue. The letter included an acknowledgement form to be returned to Elekta.
Quantity
433
Distribution
Worldwide Distribution - US Nationwide in the states of AL, CA, DC, NE, NY, ND, OR, PA, RI, SC, TX, and the countries of Angola, Argentina, Armenia, Australia, Austria, Barbados, Belarus, Brazil, Bulgaria, Canada, China Columbia, Curacao, Czech Republic, Denmark, Finland, France, Germany, Greece, Honduras, Hungary, Iceland, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Kenya, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Romania, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Yemen, Zambia.
Lot, serial or model codes
All software builds
510(k) numbers
["K121448"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register