Skip to the content

FDA Medical Device Recalls (openFDA)

Advanced Medical Optics, Inc.: CLARIS Cleaning and Soaking Solution, 4 fl. oz.

Recall number
Z-0987-05
Recalling firm
Advanced Medical Optics, Inc.
Product
CLARIS Cleaning and Soaking Solution, 4 fl. oz.
Status
Terminated
Initiated
2005-06-13
Posted
2005-07-13
Terminated
2010-02-17
Reason
Component mix-up of caps/tips. Absence of red tip indicates improperly that the solution is safe for application to the eye.
Root cause
Other
Action
Firm intiated recall to distributors on June 13, 2005 by letter.
Quantity
32,916
Distribution
Undetermined, assumed to be nationwide thru secondary distributors.
Lot, serial or model codes
Lot numbers: AA00785; AA00786: AA01050; AA01076
510(k) numbers
not on file
PMA numbers
["P870023"]
Product code
LPN
Firm FEI
3003843509
Firm city
Santa Ana
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register