Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien LP: AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair procedures Model ABSTACK30

Recall number
Z-0986-2024
Recalling firm
Covidien LP
Product
AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair procedures Model ABSTACK30
Status
Open, Classified
Initiated
2023-11-16
Posted
2024-02-06
Terminated
not on file
Reason
Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
Root cause
Process control
Action
Medtronic issued Urgent Medical Device Recall Letter and Customer Acknowledgement Form to each consignee via mail beginning 16 November 2023. Letter states reason for recall, health risk and action to take: Identify and quarantine all unused and non-expired affected lot N3J1860Y of the ModelABSTACK30 AbsorbaTack" Absorbable Fixation Device listed above. See attachment A forguidance on identifying potentially affected devices. "Return all unused product from the affected lot in your inventory to Medtronic. Please contactrs.covidienfeedbackcustomerservice@medtronic.com for the Return Good Authorization (RGA). "In addition, please complete and return the enclosed Customer Confirmation Form tors.gmbfcamitg@medtronic.com even if you do not have unused inventory. "Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed. If you have any questions regarding this communication, please contact your Medtronic Representative or Customer Service at 800-962-9888.
Quantity
54 units
Distribution
US Nationwide distribution in the states of CA, CO, FL, IL, TX.
Lot, serial or model codes
GTIN 20884523006527 Lot number: N3J1860Y
510(k) numbers
["K090470"]
PMA numbers
not on file
Product code
OCW
Firm FEI
1219930
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register