Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Med Sys Inc: Toshiba Aplio 50 Digital Ultrasound System, SSA-700A with version 6 software

Recall number
Z-0986-2007
Recalling firm
Toshiba American Med Sys Inc
Product
Toshiba Aplio 50 Digital Ultrasound System, SSA-700A with version 6 software
Status
Terminated
Initiated
2007-02-02
Posted
2007-06-27
Terminated
2012-04-18
Reason
When using the LV measurement function and making numerous measurements the value displayed on the image screen may on occasion not update correctly. The values that are stored in MDA (measurement display area) are correct and no issue has been reported with these values.
Root cause
Other
Action
Notification letters were sent to customers on February 2, 2007. Customers were informed of the safety issue regarding the possibility of incorrect measurement values being displayed on the active image screen while performing multiple measurements and instructed to verify actual results against the values stored in the MDA (measurement display area). Customers were also informed that a Toshiba Customer Service Engineer will contact them to schedule a software upgrade when software becomes available. If customers have any questions they were instructed to contact their Toshiba Customer Service Engineer or call (800) 421-1968 and ask for the Regulatory Affairs Department. Customers were also instructed to distribute this information to all users of the product and also with the Bio-medical Department at their facility.
Quantity
45
Distribution
Nationwide
Lot, serial or model codes
Serial numbers:  B2622263,  B2622264,  B2622265, B2622267, B2622269,  B3522364, B3522365, B3522366, B3522369, B3522370, B3522372, B3522373, B3522402, B3522403, B3522404, B3522405, B3552542, B3552557, B3552558, B3552560, B3562595, B3562599, B3582672, B3582673, B3582674, C3592771, C3592772, C3592773, C3592774, D4553251, D4553252, D4553253, D4563327, D4563329, E4593563, E4593564, F4603654, F4603655, F4603656, F4613730, F4613731, F4613732, F5513813, G5523871 & G5523875
510(k) numbers
["K022400"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register