Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Uroskop Omnia Max -fluoroscopic X-ray system Model # 10762473 Uroskop Omnia Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.

Recall number
Z-0985-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Uroskop Omnia Max -fluoroscopic X-ray system Model # 10762473 Uroskop Omnia Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Status
Terminated
Initiated
2019-02-08
Posted
not on file
Terminated
2020-05-05
Reason
The swivel arm holding the monitors may become dislodged from the carrying arm.
Root cause
Device Design
Action
The firm, Siemens, issued an "URGENT: MEDICAL DEVICE CORRECTION "on 2/8/19 to its customers. The letter is identifying the problem, risk, actions to take and corrective action. The customers were instructed to do the following: Immediately examine your inventory and quarantine product subject to recall (if applicable). or To avoid the risk until implementation of the corrective action, it is strongly recommended not to move and position monitors above the system table. As an alternative and if clinically applicable, monitors can also be used when they are left in the parking position. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner. The service organization will contact and arrange a date to perform this corrective action with Updated Instruction XP006/19/S. Contact the service organization for an earlier appointment at 1-800-888-7436.
Quantity
191
Distribution
US Nationwide Distribution
Lot, serial or model codes
Serial Numbers: 4032 4033 4036 4037 4038 4042 4043 4047 4052 4053 4054 4055 4056 4059 4060 4061 4063 4066 4069 4072 4073 4082 4085 4086 4094 4095 4096 4098 4099 4100 4101 4102 4104 4108 4111 4112 4114 4117 4118 4119 4120 4121 4124 4125 4126 4128 4129 4130 4132 4133 4134 4135 4136 4140 4141 4142 4143 4146 4300 4301 4303 4304 4305 4306 4309 4310 4312 4313 4314 4317 4318 4319 4321 4322 4325 4326 4328 4329 4330 4332 4333 4334 4500 4501 4505 4506 4509 4510 4511 4513 4516 4519 4520 4521 4528 4529 4530 4531 4532 4543 4544 4545 4548 4552 4553 4556 4557 4558 4559 4560 4561 4562 4565 4566 4568 4571 4572 4573 4574 4575 4578 4582 4585 4587 4588 4589 4590 4591 4593 4594 4597 4598 4600 4601 4602 4603 4605 4609 4613 4615 4616 4618 4622 4624 4626 4627 4629 4630 4631 4635 4636 4637 4641 4642 4645 4646 4649 4651 4652 4655 4659 4661 4662 4663 4664 5020 5021 5023 5025 5026 5027 5028 5030 5031 5033 5038 5043 5045 5051 5057 5058 5059 5060 5061 5066 5067 5069 5070 5072 5074 5078
510(k) numbers
["K173639"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register