Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare: GE Healthcare Revolution CT., multi-slice (256 detector row) CT scanner.

Recall number
Z-0985-2015
Recalling firm
GE Healthcare
Product
GE Healthcare Revolution CT., multi-slice (256 detector row) CT scanner.
Status
Terminated
Initiated
2014-12-19
Posted
2015-01-15
Terminated
2015-04-29
Reason
Potential safety issue due to a required torque check on three bolted joints of the Revolution CT scanner. No injuries or illnesses reported.
Root cause
Process control
Action
Consignees were sent on 12/19/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC # 25457 dated December 19, 2014. The letter was addressed to Hospital Administrators / Risk Managers, Managers of Radiology / Cardiology and Radiologists / Cardiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information (US phone 800-437-1171, Japan 0120-055-919, other countries contact your local GE Healthcare Service Representative).
Quantity
13 (7 US, 6 OUS)
Distribution
US: FL, WA, IL, UT, NY, CA. OUS: Canada, France, Hong Kong, Japan, Switzerland, United Arab Emirates.
Lot, serial or model codes
Mfg Lot or Serial # System ID 00000437461CN7 519646REVO1 00000437837CN8 M2322254 00000438342CN8 828210096 00000437028CN4 REV001 00000438864CN1 900010CT06 00000439199CN1 260010CT19 00000439068CN8 AE1477CT01 00000437787CN5 786596WKRIVO 00000438503CN5 206598REV1 00000438951CN6 309655REVCT 00000439029CN0 801662REV1 00000439160CN3 212241REVCT 00000439367CN4 415476MCT3
510(k) numbers
["K133705"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register