FDA Medical Device Recalls (openFDA)
Philips Healthcare Inc.: MultiDiagnost Eleva IITV (MD-EII) with Software Release: PBL 6.1.1 Product Code 708036 Product Usage: The Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/universal system. General R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Urinary system, Reproductive system, Respiratory system, Circulatory system, and Various: Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.
- Recall number
- Z-0985-2012
- Recalling firm
- Philips Healthcare Inc.
- Product
- MultiDiagnost Eleva IITV (MD-EII) with Software Release: PBL 6.1.1 Product Code 708036 Product Usage: The Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/universal system. General R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Urinary system, Reproductive system, Respiratory system, Circulatory system, and Various: Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.
- Status
- Terminated
- Initiated
- 2012-01-09
- Posted
- 2012-02-08
- Terminated
- 2017-02-13
- Reason
- Incorrect Skin Dose calculation. AKR (Air Kerma Rate) displayed and stored in patient records is too low
- Root cause
- Employee error
- Action
- Philips Healthcare issued a mandatory Urgent Field Safety Noticedated October 3, 2011 to all affected customers. The letter identifies the affected product, problem description, hazard involved, action to be taken by customer/user and actions planned by Philips to correct the problem. The letter provide important information for the continued safe and proper use of the equipment. The letter instructs customers to review the following information with all members of their staff who need to be aware of the contents of this communication and retain a copy with the equipment Instruction for Use. For further information or support concerning this issue contact your local Philips representative at 866-767-2822.
- Quantity
- 10 units (US)
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of CA, IL, OH, SD, TX, WY and the countries of: Australia, Austria, Canada, Egypt, Germany, mexico, Turkey and UK
- Lot, serial or model codes
- Software Release: PBL 6.1.1
- 510(k) numbers
- ["K023441"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28