Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Inc.: MultiDiagnost Eleva IITV (MD-EII) with Software Release: PBL 6.1.1 Product Code 708036 Product Usage: The Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/universal system. General R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Urinary system, Reproductive system, Respiratory system, Circulatory system, and Various: Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.

Recall number
Z-0985-2012
Recalling firm
Philips Healthcare Inc.
Product
MultiDiagnost Eleva IITV (MD-EII) with Software Release: PBL 6.1.1 Product Code 708036 Product Usage: The Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/universal system. General R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Urinary system, Reproductive system, Respiratory system, Circulatory system, and Various: Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.
Status
Terminated
Initiated
2012-01-09
Posted
2012-02-08
Terminated
2017-02-13
Reason
Incorrect Skin Dose calculation. AKR (Air Kerma Rate) displayed and stored in patient records is too low
Root cause
Employee error
Action
Philips Healthcare issued a mandatory Urgent Field Safety Noticedated October 3, 2011 to all affected customers. The letter identifies the affected product, problem description, hazard involved, action to be taken by customer/user and actions planned by Philips to correct the problem. The letter provide important information for the continued safe and proper use of the equipment. The letter instructs customers to review the following information with all members of their staff who need to be aware of the contents of this communication and retain a copy with the equipment Instruction for Use. For further information or support concerning this issue contact your local Philips representative at 866-767-2822.
Quantity
10 units (US)
Distribution
Worldwide Distribution - USA (nationwide) including the states of CA, IL, OH, SD, TX, WY and the countries of: Australia, Austria, Canada, Egypt, Germany, mexico, Turkey and UK
Lot, serial or model codes
Software Release: PBL 6.1.1
510(k) numbers
["K023441"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register