Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: Dual Articulating Headrest, Velcro P/N P141210813, REF BF753

Recall number
Z-0984-2024
Recalling firm
Steris Corporation
Product
Dual Articulating Headrest, Velcro P/N P141210813, REF BF753
Status
Open, Classified
Initiated
2023-12-12
Posted
2024-02-05
Terminated
not on file
Reason
Dual-articulating head rest may not stay in place or may fail to lock into position during a patient procedure which could result in an injury to the patient or a procedural delay.
Root cause
Process control
Action
An Urgent Medical Device Recall notice dated December 5, 2023 was issued to distributors and customers via FedEx explaining the removal of specific affected product . The notice explains the labeled date of manufacture must fall between August 14, 2023 and October 26, 2023 the recall notice must have been received directly from STERIS. The Dual-Articulating Head Rest may not stay in place or may fail to lock into position during a patient procedure which could result in an injury to the patient or a procedural delay. STERIS Service Technician will work with your facility to schedule a time to hand-deliver replacement head rest(s) and pick up your affected head rest(s). It is recommended all head rests impacted by this recall be immediately removed from service until it can be replaced. If you have further distributed this product, please identify all locations, notify them of this recall, and coordinate the collection of all affected head rest(s).
Quantity
80
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, IA, KS, LA, MN, MS, NV, NJ, OH, OK, OR, PA, TN, TX, UT, WA, WI & Puerto Rico, and the countries of Australia, Canada, Mexico & New Zealand.
Lot, serial or model codes
UDI-DI: 00724995195885, date of manufacture 2023-08-14 thru 2023-10-26.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HBM
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register