Beckman Coulter, Inc.: Beckman Coulter IRISpec CA/CB/CC control, Catalog #800-7211, containing 3 x 100 mL bottles each of CA Control, CB Control, and CC Control.

Recall number
Z-0984-2022
Recalling firm
Beckman Coulter, Inc.
Product
Beckman Coulter IRISpec CA/CB/CC control, Catalog #800-7211, containing 3 x 100 mL bottles each of CA Control, CB Control, and CC Control.
Status
Open, Classified
Initiated
2022-02-16
Posted
not on file
Terminated
not on file
Reason
IRISpec CA failures for glucose.
Root cause
Under Investigation by firm
Action
The recall letter was issued via email on 2/17/2022 and by first class mail on 2/18/2022 for U.S. consignees. The letter explained the issue, the impact of the issue, and the actions to be taken by the consignee. The consignee was to update the quality control QC ranges for glucose as instructed in the letter. The ranges superseded those listed in the IFU and the assay sheet C39721AA, and the letter emphasized to use only the specifications in the notification. As an interim action, a lower limit on the glucose control range was being implemented via the notification. The letter was to serve as a labeling update for the issue until further notice. The information was to be shared with the consignee's laboratory staff and retained as part of the laboratory Quality System documentation. If the product has been forwarded to another laboratory, the consignee is to provide them a copy of the letter. The consignee was requested to respond to the notification electronically or manually within 10 days. A response form was enclosed for completion and return. Also enclosed was an addendum which provided instructions for updating the QC ranges for glucose in the analyzer. OUS consignees were notified via email and first class USPS beginning 2/16/2022.
Quantity
12,306 kits
Distribution
Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Brazil, Canada, Curacao, Guatemala, Mexico, Philippines, Singapore, and Taiwan.
Lot, serial or model codes
Lot #583-21 and higher, UDI 10837461002611.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJW
Firm FEI
3000203293
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register