Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems North America Co. Phillips: Panorama 1.0T MRI

Recall number
Z-0984-2007
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Panorama 1.0T MRI
Status
Terminated
Initiated
2007-04-30
Posted
2007-06-27
Terminated
2011-09-21
Reason
Potential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site.
Root cause
Other
Action
On 4/30/07 Philips sent an Urgent Device Correction letter dated April 25, 2007 via certified mail to customers. The letter advises customers of the issue and that that a representative will visit the sites, inspect and correct the problem if indicated.
Quantity
56 units
Distribution
Units were distributed throughout the US to 1,078 hospitals and medical centers.
Lot, serial or model codes
Site #s: 521810, 533745, 533860, 533900, 534276, 534278, 535230, 535273, 535664, 536972, 537852, 538574, 538616, 539265, 539374, 539514, 540482, 541065, 541851, 542236, 542244, 543822, 544039, 544148, 544696, 544723, 545034, 545143, 545732, 545902, 545918, 545981, 546206, 546594, 546654, 546722, 546728, 547843, 547897, 548040, 548041, 548064, 549389, 549909, 550324, 550358, 550398, 550439, 550449, 550621, 551025, 551583, 552453, 552526, 552609 and 47480028.
510(k) numbers
["K041602"]
PMA numbers
not on file
Product code
LNH
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register