FDA Medical Device Recalls (openFDA)

Abiomed, Inc.: Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 1000080 and 0048-0045 - Impella CP with SmartAssist ***Updated February 2024*** (10) 004413 - Impella 2.5 Set (12) 0046-0026 - Impella 5.0 Pump Set ROW (13) 0046-0037 - Impella 5.0 Pump Set APAC (15) 004680-AU - 5.0 Pump Set AU (16) 0048-0002 - Impella CP Pump Set, EU (17) 0048-0002-BR - Impella CP Pump Set BR (18) 0048-0004 - Impella CP Pump Set, Canada (19) 0048-0014 - Impella CP Smart Assist Set, EU (20) 0048-0024-JP - Impella CP Smart Assist Set, JP (21) 0048-0044 - Impella CP Smart Assist Set, Canada (22) 0048-0047 - Impella CP Smart Assist Set APAC (23) 005040 - Impella 2.5 IMC Pump Set EU (24) 005048-JP - Impella 2.5 Pump Set, Japan (25) 005060 - Impella 5.0 IMC Pump Set EU (26) 005064 - Impella 5.0 IMC Pump Set Canada (27) 005066-JP - Impella 5.0 Pump Set, Japan (28) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (29) 0550-0004 - Impella 5.5 with SmartAssist Set, CA (30) 1000115 - Impella CP Pump set, APAC (31) 1000211 - Impella 5.5 SmartAssist Set, JP (32) 1000302 - Impella CP with SmartAssist APAC (33) 1000361 - Impella 5.5 Set AU (34) 1000402 - Impella CP Smart Assist Set ***Updated March 15, 2024*** The following products were removed from the list of affected products because the Right-side devices are not in the scope of the warning to avoid left ventricle perforation: (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (11) 0046-0011 - Impella RP Pump Set, EU (14) 0046-0039 - Impella RP Set APAC

Recall number
Z-0980-2024
Recalling firm
Abiomed, Inc.
Product
Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 1000080 and 0048-0045 - Impella CP with SmartAssist ***Updated February 2024*** (10) 004413 - Impella 2.5 Set (12) 0046-0026 - Impella 5.0 Pump Set ROW (13) 0046-0037 - Impella 5.0 Pump Set APAC (15) 004680-AU - 5.0 Pump Set AU (16) 0048-0002 - Impella CP Pump Set, EU (17) 0048-0002-BR - Impella CP Pump Set BR (18) 0048-0004 - Impella CP Pump Set, Canada (19) 0048-0014 - Impella CP Smart Assist Set, EU (20) 0048-0024-JP - Impella CP Smart Assist Set, JP (21) 0048-0044 - Impella CP Smart Assist Set, Canada (22) 0048-0047 - Impella CP Smart Assist Set APAC (23) 005040 - Impella 2.5 IMC Pump Set EU (24) 005048-JP - Impella 2.5 Pump Set, Japan (25) 005060 - Impella 5.0 IMC Pump Set EU (26) 005064 - Impella 5.0 IMC Pump Set Canada (27) 005066-JP - Impella 5.0 Pump Set, Japan (28) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (29) 0550-0004 - Impella 5.5 with SmartAssist Set, CA (30) 1000115 - Impella CP Pump set, APAC (31) 1000211 - Impella 5.5 SmartAssist Set, JP (32) 1000302 - Impella CP with SmartAssist APAC (33) 1000361 - Impella 5.5 Set AU (34) 1000402 - Impella CP Smart Assist Set ***Updated March 15, 2024*** The following products were removed from the list of affected products because the Right-side devices are not in the scope of the warning to avoid left ventricle perforation: (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (11) 0046-0011 - Impella RP Pump Set, EU (14) 0046-0039 - Impella RP Set APAC
Status
Open, Classified
Initiated
2023-12-27
Posted
2024-02-09
Terminated
not on file
Reason
IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
Root cause
Labeling Change Control
Action
Firm notified affected consignees of the updated warnings in the IFU beginning on December 27, 2023. Customers were provided with details of the IFU modifications. Customers should notify everyone at their facility who needs to be informed of this correction. If products have been forwarded to another facility, please notify that facility of the updated IFUs. Post a copy of the recall notice in a visible area for awareness. If you have questions or concerns regarding this notice, please contact recallcoordinators@abiomed.com and/or your local clinical field staff.
Quantity
91,914 (65,857 US; 26,075 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.
Lot, serial or model codes
UDI-DI (1) 00813502011081 (2) 00813502011180 (3) 00813502011227 (7) 00813502011531 and 00813502012828 (8) 00813502011388 (9) 00813502012279 and 00813502011876 All product IFUs include the Update ***Added February 2024*** (10) 813502010947 (12) 813502011821 (13) 813502011937 (15) N/A (16) 4260113630242 (17) 813502011777 (18) 813502011265 (19) 4260113630280 (20) 813502011609 (21) 813502012200 (22) 813502011944 (23) 4260113630136 (24) 813502010046 (25) 4260113630174 (26) 813502011197 (27) 813502010053 (28) 813502011630 (29) 813502010466 (30) 813502010473 (31) 813502012453 (32) 813502012767 (33) 813502012873 (34) 813502012873 ***Removed from scope March 15, 2024*** (4) 00813502011029 (5) 00813502011869 (6) 00813502012811 (11) 4260113630273 (14) 813502011951
510(k) numbers
not on file
PMA numbers
["P140003","P170011"]
Product code
OZD
Firm FEI
1220648
Firm city
Danvers
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register