Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: 133" 60 Drop 24 Hour Primary Set w/2 Bravo24, 2 Gang 1o2 Manifold, Rotating Luer, 1 Ext, Item No. B59006 Product Usage: The intravascular administration set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein). Each set will be manufactured to a unique specifications using dimensions, components and configurations specified by the customer.

Recall number
Z-0980-2017
Recalling firm
ICU Medical, Inc.
Product
133" 60 Drop 24 Hour Primary Set w/2 Bravo24, 2 Gang 1o2 Manifold, Rotating Luer, 1 Ext, Item No. B59006 Product Usage: The intravascular administration set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein). Each set will be manufactured to a unique specifications using dimensions, components and configurations specified by the customer.
Status
Terminated
Initiated
2016-11-30
Posted
2017-01-10
Terminated
2017-06-23
Reason
ICU Medical Inc. has identified an isolated event involving a 15 drop drip chamber assembled on a device which should contain a 60 drop drip chamber.
Root cause
Mixed-up of materials/components
Action
An Urgent Medical Device Recall Notification letter dated 11/30/16 was sent to their customers to inform them that ICU Medical Inc. has identified an isolated event involving a 15 drop drip chamber assembled on a device which should contain a 60 drop drip chamber. The letter instructs the customers of the actions required and to return completed recall response form to ICU Medical via fax (801) 264-1755 or by email recall@icumed.com. Customers with questions are instructed to contact ICU Customer Service Monday through Friday between 8:30AM and 4PM Pacific Time, (866) 829-9025 and select option 2 or email productreturns@icumed.com.
Quantity
225 units
Distribution
US: NC
Lot, serial or model codes
Lot No. 3323650
510(k) numbers
K964435
PMA numbers
not on file
Product code
FMG
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source