Skip to the content

FDA Medical Device Recalls (openFDA)

Lifescan Inc: LifeScan brand One Touch UltraSmart Meter Owners Booklet AW 060-788-01A; All One Touch UltraSmart Meters shipped with OneTouch UltraSmart Owner''s Booklet AW 060-788-01A are affected by this Field Correction.

Recall number
Z-0980-05
Recalling firm
Lifescan Inc
Product
LifeScan brand One Touch UltraSmart Meter Owners Booklet AW 060-788-01A; All One Touch UltraSmart Meters shipped with OneTouch UltraSmart Owner''s Booklet AW 060-788-01A are affected by this Field Correction.
Status
Terminated
Initiated
2005-06-09
Posted
2005-07-13
Terminated
2006-12-21
Reason
Meter labeling (Owner's Booklet and Quick Reference Guide) is incorrect as it indicates that the meter displays the message 'Do you need a snack' for glucose values between 20 and 70 mg/dL; whereas the OneTouch UltraSmart meter displays the message, 'Do you need a snack?' for glucose values from 20 through 59 mg/dL.
Root cause
Other
Action
On 6/9/05, the firm initiated the recall and their notification was via letters informing its customers of the corrective action.
Quantity
584,291 units
Distribution
Product was released for distribution to 40,763 consignees worldwide. The firm provided the following distribution information: 124 wholesaler dealers, 224 distributors, 140 retailers and 40,275 consumers/users involved in the recall. The recall product was distributed in all States within the U.S. In addition, OneTouch UltraSmart meter kits with the affected UltraSmart Owner''s Booklets (AW 060-788-01A) have been distributed internationally in Canada, France, Germany, Italy, United Kingdom, Benelux, Sweden, Finland, Norway, Denmark, Spain, Portugal, Switzerland and Austria. The recall was appropriately extended to the consumer/user level; i.e., the wholesalers/distributors, pharmacists, and patients who received the recalled products.
Lot, serial or model codes
All One Touch UltraSmart Meters shipped with OneTouch UltraSmart Owner''s Booklet AW 060-788-01A are affected by this Field Correction.  Part Number: Description: 020-524-01, System,OUS,US; 020-524-01F, System,OUS,US; 020-524-01Y, UltraSmart Meter; 020-524-02, System,OUS,EAS-C,US; 020-524-03, System,OUS,EAS-S,US; 020-524-04F, OT UltraSmart System Kit V. 4 USA; 020-524-04Y, UltraSmart System; 020-524-05, UltraSmart System -EAS-C; 020-524-06, UltraSmart System -EAS-S; 020-524-10, OT UltraSmart System Kit; 020-524-10F, OT UltraSmart System Kit;  OT UltraSmart System Kit V. 10 USA; 020-524-10Y, OT UltraSmart System Kit; 020-524-11, System, EAS-C, OUS, US; 020-524-12, System, EAS-S, OUS, US; 020-524-13, OT UltraSmart System; 020-759-02F, OT UltraSmart Meter V.2 USA; 020-759-03F, OT UltraSmart Meter V.3 USA; 020-761-01F, Starter,OUS,US; 020-761-02, OT UltraSmart Starter; 020-761-02F, OT UltraSmart Starter; 020-761-04, OT UltraSmart Starter; 020-761-04F OT UltraSmart Starter; 020-761-05 OT UltraSmart Starter Kit; 020-761-05F, OT UltraSmart Starter Kit; 020-926-01, Dist Kit,SM,OTU,US; 020-930-01, Dist,Kit,OTU/OUS,US; 020-971-01, Disp,Counter,Ultra Kits; 021-075-01, UltraSmart SVP; 021-113-01, Display, Hanging, OUT/OUS, US; 021-133-01, Display, Counter, OTU/OUS Kits; 021-135-01, Span Display, Hanging, OTU, US; and 021-166-01, Display, Hanging, OTU/OUS Kits.
510(k) numbers
["K021819"]
PMA numbers
not on file
Product code
NBW
Firm FEI
2939301
Firm city
Milpitas
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register