Philips Medical Systems (Cleveland) Inc: Various Brilliance and Ingenuity Computed Tomography X-ray systems; one unit per package Product Usage: Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
- Recall number
- Z-0979-2015
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Various Brilliance and Ingenuity Computed Tomography X-ray systems; one unit per package Product Usage: Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
- Status
- Terminated
- Initiated
- 2014-12-08
- Posted
- 2015-01-15
- Terminated
- 2015-10-01
- Reason
- Improper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.
- Root cause
- Process control
- Action
- Philips sent an Urgent - Field Safety Notice letter dated December 4, 2014 to all affected customers. The US consignees will receive this notice via certified mail and Philips Key Market representatives will distribute the FSN internationally. Additionally, A Field Service Engineer will service the system to resolve the issue. For questions contact your local Philips representative or local Philips Healthcare office.
- Quantity
- 500 units
- Distribution
- US Nationwide Distribution - AR, AZ, CA, CO, FL, FL, GA, IL, IN, MD, MI, MN, NC, ND, NJ, NY, OH, OK, OR, PA, PA, SC, TN, TX, VA, and WA. The affected product was also distributed to one government consignee. The affected product was also distributed to the following foreign countries: Canada, Albania, Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Cayman Islands, China, Czech Republic, Denmark, Ecuador, Ethiopia, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Libyan Arab Jamahiriya, Malaysia, Mauritius, Mexico, Myanmar, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, United Kingdom, and Vietnam.
- Lot, serial or model codes
- Brilliance CT 16 Air: Model Number 728246: Serial Numbers: 3117,6680,6692,6693,6697,6698,6700,6703,6706,6707,6708,6709,6711,6712,6713, 6715,6716,6718,6719,6720,6721,6722,6723,6724,6726,6727,6728,6729,6730,6731, 6732,6734,6735,6736,6737,6738,6740,6742,6743,6744,6746,6747,6748,6749,6750, 6751,6752,6753,6754,6755,6756,6757,6759,6760,6761,6762,6764,6765,6766,6767, 6769,6770,6771,6772,6774,6775,6778,6779,6783,6784,50209,50213,50215,50216, 50217,50221,50222,50223,50224,50225,50226,50227,260013,260015,260016, 260017,260018,260025,260027,260028,260029,260030,260032,260035,260042, 260043,260045,260046,270002,270003,270004,270005,270006 Brilliance CT 64: Model Number 728231: Serial Numbers: 10773,10793,10795,10798,10803,10805,95061,95857,95861,95865,95880,95881, 95882,95883,95885,95892,95895,95900,95908,95909,95910,95912,95915,95916, 95918,95920,95921,95922,95924,95925,95926,95928,95930,95933,95936,95944 Brilliance CT Big Bore Oncology: Model Number 728243: Serial Numbers: 7884,7890,7909,7923,7925,7927,7928,7929,7936,7938,7940,7941,7942,7944,7945, 7948,7949,7955,7956,7963,7968,7969,7970,7971,7973,7975,7976,7977,7980,7984, 7989,7990,7991,7994,7995,7996,7999,8002,8005,8006 Brilliance CT Big Bore Radiology: 728244 Serial Numbers: 7879,7897,7912,7919,7922,7926,7930,7943,7957,7958,7959,7967,7979,7985,7988, 7992 Brilliance iCT: Model Number 728306: Serial Numbers: 10065,100428,100432,100445,100450,100451,100452,100454,100455,100460, 100463,100465,100466,100470,100471,100474,100475,100476,100477,100478, 100479,100480,100481,100483,100485,100486,100487,100488,100489,100490, 100491,100492,100493,100494,100496,100498,100499,100500,100502,100503, 100504,100505,100507,100508,100509,100510,100514,100515,100517,100520 Brilliance iCT SP: Model Number 728311: Serial Numbers: 200126,200131,200137,200138 Ingenuity CT: Model Number 728326: Serial Numbers: 30005,30006,30009,30011,30012,30013,30014,30015,91007,300022,300028,300032,300035,300036,300156,300157,300158,300161,300162,300164,300165,300166, 310170 Ingenuity Core: Model Number 728321: Serial Numbers: 31001,52001,52002,52003,52004,52006,52007,52009,52010,52011,52013,52015, 52016,52017,52018,52020,52021,52024,52026,52030,52031,52032,52033,310032, 310090,310113,310114,310115,310116,310121,310129,310131,310134,310135, 310137,310138,310139,310140,310144,310145,310147,310148,310149,310151, 310152,310153,310154,310155,310158,310160,310161,310162,310163,310164, 310165,310170,310171,333005,333009,333011,333012,333013,333014,333015, 333016,333017,333020,333024,333026,333031,Unknown,Unknown Ingenuity Core128: Model Number 728323: 30153,32001,32002,32003,32004,32005,32006,32007,32009,32010,32011,32012, 32013,32014,32015,32016,32018,32020,32021,32023,32024,32025,32026,32029, 32030,32031,32032,32033,32034,32035,32036,32037,32038,32040,32041,32042, 32043,32044,32046,32047,32048,32049,32051,32052,32054,32062,32064,32065, 32066,32067,32068,32069,32070,320022,320097,320101,320102,320105,320106, 320107,320109,320112,320113,320114,320116,320117,320118,320120,320121, 320122,320124,320125,320126,320128,320129,320131,320134,320137,320138, 320139,320140,320141,320142,320143,320144,320145,320146,320147,320148, 320150,336001,336002,336006,336007,336008,336009,336010,336011,336012, 336013,336014,336015,336016,336018,336020,336021,336024,336025,336026, 336027,336028,336029,336030,336031,336032,336033,336034,336035,336036, 336037,336038,336039,336040,336042,336043,336044,336046,336047,336048, 336050,336051,336052,336055,336056,336057,336064,336072 Ingenuity Flex: Model Number 728317: 260002,260003,340002,345001,345002,345003,345004,345005,345006,345007, 345010,345012,345013,345014,345016,345017,345018,345022,345024
- 510(k) numbers
- ["K012009","K033326","K033357","K060937"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28