Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet Spine, LLC: Timberline Anchored Lateral Variable Retractable Sleeve LITPCR8600-2101-07, Model No. 86300204 Timberline Anchored Lateral Variable Retractable Sleeves (the Sleeves) are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.

Recall number
Z-0978-2015
Recalling firm
Biomet Spine, LLC
Product
Timberline Anchored Lateral Variable Retractable Sleeve LITPCR8600-2101-07, Model No. 86300204 Timberline Anchored Lateral Variable Retractable Sleeves (the Sleeves) are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.
Status
Terminated
Initiated
2013-08-22
Posted
2015-01-14
Terminated
2015-03-13
Reason
The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve shaft.
Root cause
Component design/selection
Action
The firm, Lanx, sent their Consignees/customers a letter dated October 15, 2013. The letter described the product, problem and actions taken. Lanx and their distributors replaced the affected inventory with the conforming product. Please contact the Chief Compliance Officer at 303-501-8534 or by email: Randy.fesmire@lanx.com.
Quantity
12 units
Distribution
US Distribution to states of: AZ, CA, NY, TN, and TX including PR.
Lot, serial or model codes
Model 8630-0204 Lots TU00052, TU00052A, and TU00117
510(k) numbers
["K123767"]
PMA numbers
not on file
Product code
OVD
Firm FEI
3012447612
Firm city
Broomfield
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register