Skip to the content

FDA Medical Device Recalls (openFDA)

King Systems Corp.: KING LTS -D, Size 4, 10-0026, Oropharyngeal Airway. King Systems Corporation 15011 Herriman Boulevard, Noblesville, IN 46060. Intended use: Airway management by providing a patent airway to allow patient ventilation.

Recall number
Z-0978-2011
Recalling firm
King Systems Corp.
Product
KING LTS -D, Size 4, 10-0026, Oropharyngeal Airway. King Systems Corporation 15011 Herriman Boulevard, Noblesville, IN 46060. Intended use: Airway management by providing a patent airway to allow patient ventilation.
Status
Terminated
Initiated
2009-12-15
Posted
2011-01-27
Terminated
2012-08-16
Reason
King Systems was issued a warning letter from the Center for Devices and Radiological Health, Office of Compliance on October 28, 2009. The warning letter asserts that Kings Systems KLT(S)-D airway exceeds cleared indications in that it is promoted for "airway management in patients over 4 feet in height, for controlled or spontaneous ventilation, difficult and emergent airway cases, and ambulat
Root cause
Labeling False and Misleading
Action
On October 29, 2009, the firm, Kings System, sent a "Urgent - Product Correction" letter to the customers. The letter described the product, problem and actions to be taken. The customers were instructed to remove the promotional material from the King System website which includes literature, and Instructions for Use. The firm instructed all Sales force and distributors to cease dissemination of promotional materials, literature and communications. Note: The firm revised indications, instructions for use, product labels and labeling so it is aligned with the 510Ks. The customers were also instructed to replace current instructions for use with revised instructions for use and destroy the current IFU; return Kings LT(S)-D oropharyngeal airway product inventory to King Systems; and complete and return the attached Instructions for Use Destruction Signature Form via fax to King Systems at 317-776-5197. For further information, please call 317-776-6823 ext 257 or email: LTSD@kingsystem.com.
Quantity
25,000 all products.
Distribution
Worldwide distribution: USA and countries including: Canada, Costa Rica, and Panama.
Lot, serial or model codes
all product produced prior to 12/2009.
510(k) numbers
["K021634","K033186"]
PMA numbers
not on file
Product code
CAE
Firm FEI
1824226
Firm city
Noblesville
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register