Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810 Implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

Recall number
Z-0977-2025
Recalling firm
Integra LifeSciences Corp.
Product
CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810 Implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.
Status
Open, Classified
Initiated
2024-12-16
Posted
2025-01-17
Terminated
not on file
Reason
Potential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.
Root cause
Labeling Change Control
Action
Integra LifeSciences notified consignees on about 12/16/2024 via letter. Consignees were instructed to review and understand the information provided in the letter, complete and return the provided Customer Acknowledgement Form, remove any affected units from service and quarantine them, arrange for return of affected units, and forward the notification to anyone who utilizes the affected units. Distributors were instructed to follow the same directions but also requested to notify any customers if product was further distributed, and return affected units.
Quantity
132 units
Distribution
US Nationwide distribution in the states of FL, NC, MA, MN, OH, DC, VA, CA, WI, NE, TX, NY, MI, OR, NJ, SC, AZ, UT, RI, IN, GA, CO and the countries of Canada, Australia, New Zealand, Cyprus, Sweden, Finland, Slovenia and Great Britain.
Lot, serial or model codes
UDI-DI: 10381780535874. Lot Numbers: 7432470, 7434917, 7434918 and 7438322.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JXG
Firm FEI
3003418325
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register